device · product 1 of 1
Clinical Chemistry, ABX Pentra Lipase CP, ABX Pentra 400/Pentrac400, REF: A11A01631, Model/Part # 1220001631, UDI # 03610230004944
- Recall number
- Z-0119-2020
- Initiated
- October 30, 2018
- Classification
- Class II
- Status
- Terminated
- Recalling firm
- Horiba Instruments Incorporated
- Quantity
- 1539 Chemistry Analyzers
App-derived interpretation
Potential for chemical analyzer to report abnormal results due to an interference with triglycerides CP and Cholesterol CP. An over estimation of the lipase concentration could be observed and a patient result could be declared abnormal by the instrument due to these interferences instead of normal result < 38 U/L (or < reference value used in your laboratory).
Official device-enrichment evidence · Sourced
Device Design
Inspect official wording and provenance
Reason for recall
Potential for chemical analyzer to report abnormal results due to an interference with triglycerides CP and Cholesterol CP. An over estimation of the lipase concentration could be observed and a patient result could be declared abnormal by the instrument due to these interferences instead of normal result < 38 U/L (or < reference value used in your laboratory).
Code information
All Lots; All Software versions
Distribution pattern
US: AL, AZ, CA, CO, CT, FL, GA, IA, ID, IL, KS, KY, LA, MA, MD, MI, MN, MO, MS, MT, NC, NE, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, TN, TX, UT, VA, WA, WI, WV, WY