Recall Observatory FDA recall evidence

Exact FDA event identifier

Event 83398

4 recalled-product records grouped only because every row carries this exact official event ID.

Open FDA event record

Event summary

Timeline bucket
January 28, 2019
Product types
Device
Classifications
Class II
Statuses
Ongoing
Recalling firm wording
Siemens Medical Solutions USA, Inc.

Dossier provenance

Source snapshots represented here

  • openFDA Device Enforcement Report-date coverage June 20, 2012–July 08, 2026 Known gap: openFDA describes 2004-present coverage, but the current export has no report-date records in this interval.
  • openFDA Device Recall event_date_initiated coverage June 01, 1997–June 30, 2026

Enforcement coverage uses report dates; product initiation dates can precede those bounds. Device-enrichment coverage uses FDA event_date_initiated.

How coverage and grouping work

Complete imported group

Every recalled product in this event

4 official enforcement rows

Product dates, classifications, firms, and source wording remain attached to their individual rows. Opening this dossier never hides products that did not match a prior timeline filter.

device · product 1 of 4

ACUSON NX2 Diagnostic Ultrasound System, Model # 11284381, RX Only, Siemens Medical Solutions, containing software version VA10A, VA10B, VA11A, VA11B; UDI # 04056869030005 The ACUSON NX2 diagnostic ultrasound imaging systems are intended for the following applications: Cardiac, Fetal, Abdominal (including liver), Pediatric, Small Parts (Small Organ), Transcranial, OB/GYN, Pelvic, Urology, Vascular (including Peripheral Vessel) and Musculoskeletal applications. The systems also provide for the measurement of anatomical structures and for analysis packages that provide information that is used for clinical diagnosis purposes.

Z-2350-2019
Recall number
Z-2350-2019
Initiated
January 28, 2019
Classification
Class II
Status
Ongoing
Quantity
1869 systems

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
The action is being initiated due to internal testing which identified a possibility for transducers to exceed the acoustic output power (AOP) values defined . The transit voltage values based on PW Doppler are calculated normally but not loaded correctly into hardware

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Sourced

Software design

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

The action is being initiated due to internal testing which identified a possibility for transducers to exceed the acoustic output power (AOP) values defined . The transit voltage values based on PW Doppler are calculated normally but not loaded correctly into hardware

Code information

All product lots of the Acuson NX2, model # 11284381 containing the following software: VA10A, VA10B,VA10C, VA11A, VA11B, VA11C

Distribution pattern

Worldwide Distribution: US (nationwide) including states of: AL, AZ, AR, CA, CO, CT, DC, FL, GA, IL,IA, IN, KY, LA, MD, MA, MI, MN, MS, NE, NV, NH, NJ, NM, NY, NC, ND, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VT, VA, WA, WV, WI and WY; and countries of: Albania, Algeria, Angola, Argentina, Australia, Austria, Azerbaijan, Bahrain, Bangladesh, Belarus, Belgium, Bolivia, Bosnia, Brazil, Brunei, Bulgaria, Burkina Faso, Cameroon, Canada, Chile, China, Colombia, Costa Rica, Croatia, Cyrus, Czech Republic, Democratic Republic of Congo, Denmark, Dominican Republic, Ecuador, Egypt, El Salvador Finland, France, Gabon, Georgia, Germany, Greece, Guatemala, Hong Kong, Hungary, India, Indonesia, Iran, Iraq, Ireland, Israel, Italy, Ivory Coast, Japan, Jordan, Kazakhstan, Kenya, Kosovo, Latvia, Lebanon, Luxembourg, Macedonia, Malaysia, Maldives, Mali, Mexico, Morocco, Nepal, Netherlands, Norway, Pakistan, Palestine, Panama, Philippines, Poland, Portugal, Republic of Korea, Romania, Ruanda, Russian Federation, Saudi Arabia, Senegal, Serbia, Singapore, Slovakia, Slovenia, South Africa, Spain, Sri Lanka, Sweden, Switzerland, Taiwan, Tanzania, Thailand, Turkey, U.A.E., Uganda, Ukraine, United Kingdom, Uruguay, Uzbekistan and Vietnam.

device · product 2 of 4

ACUSON NX2 Elite Diagnostic Ultrasound System, Model # 11284500, RX Only, Siemens Medical Solutions, containing software version VA10A, VA10B, VA11A, VA11B; UDI # 04056869050775 The ACUSON NX2 diagnostic ultrasound imaging systems are intended for the following applications: Cardiac, Fetal, Abdominal (including liver), Pediatric, Small Parts (Small Organ), Transcranial, OB/GYN, Pelvic, Urology, Vascular (including Peripheral Vessel) and Musculoskeletal applications. The systems also provide for the measurement of anatomical structures and for analysis packages that provide information that is used for clinical diagnosis purposes.

Z-2351-2019
Recall number
Z-2351-2019
Initiated
January 28, 2019
Classification
Class II
Status
Ongoing
Quantity
70 systems

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
The action is being initiated due to internal testing which identified a possibility for transducers to exceed the acoustic output power (AOP) values defined . The transit voltage values based on PW Doppler are calculated normally but not loaded correctly into hardware

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Sourced

Software design

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

The action is being initiated due to internal testing which identified a possibility for transducers to exceed the acoustic output power (AOP) values defined . The transit voltage values based on PW Doppler are calculated normally but not loaded correctly into hardware

Code information

All product lots of the Acuson NX2, model # 11284500 containing the following software: VA10A, VA10B,VA10C, VA11A, VA11B, VA11C

Distribution pattern

Worldwide Distribution: US (nationwide) including states of: AL, AZ, AR, CA, CO, CT, DC, FL, GA, IL,IA, IN, KY, LA, MD, MA, MI, MN, MS, NE, NV, NH, NJ, NM, NY, NC, ND, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VT, VA, WA, WV, WI and WY; and countries of: Albania, Algeria, Angola, Argentina, Australia, Austria, Azerbaijan, Bahrain, Bangladesh, Belarus, Belgium, Bolivia, Bosnia, Brazil, Brunei, Bulgaria, Burkina Faso, Cameroon, Canada, Chile, China, Colombia, Costa Rica, Croatia, Cyrus, Czech Republic, Democratic Republic of Congo, Denmark, Dominican Republic, Ecuador, Egypt, El Salvador Finland, France, Gabon, Georgia, Germany, Greece, Guatemala, Hong Kong, Hungary, India, Indonesia, Iran, Iraq, Ireland, Israel, Italy, Ivory Coast, Japan, Jordan, Kazakhstan, Kenya, Kosovo, Latvia, Lebanon, Luxembourg, Macedonia, Malaysia, Maldives, Mali, Mexico, Morocco, Nepal, Netherlands, Norway, Pakistan, Palestine, Panama, Philippines, Poland, Portugal, Republic of Korea, Romania, Ruanda, Russian Federation, Saudi Arabia, Senegal, Serbia, Singapore, Slovakia, Slovenia, South Africa, Spain, Sri Lanka, Sweden, Switzerland, Taiwan, Tanzania, Thailand, Turkey, U.A.E., Uganda, Ukraine, United Kingdom, Uruguay, Uzbekistan and Vietnam.

device · product 3 of 4

ACUSON NX3 Diagnostic Ultrasound System, Model # 11235612, RX Only, Siemens Medical Solutions, containing software versions: VA10A, VA10B, VA10C, VA10D, VA10E, VA10F, VB20A, and VB20B; UDI # 04056869004297 The ACUSON NX2 diagnostic ultrasound imaging systems are intended for the following applications: Cardiac, Fetal, Abdominal (including liver), Pediatric, Small Parts (Small Organ), Transcranial, OB/GYN, Pelvic, Urology, Vascular (including Peripheral Vessel) and Musculoskeletal applications. The systems also provide for the measurement of anatomical structures and for analysis packages that provide information that is used for clinical diagnosis purposes.

Z-2352-2019
Recall number
Z-2352-2019
Initiated
January 28, 2019
Classification
Class II
Status
Ongoing
Quantity
1384 systems

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
The action is being initiated due to internal testing which identified a possibility for transducers to exceed the acoustic output power (AOP) values defined . The transit voltage values based on PW Doppler are calculated normally but not loaded correctly into hardware

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Sourced

Software design

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

The action is being initiated due to internal testing which identified a possibility for transducers to exceed the acoustic output power (AOP) values defined . The transit voltage values based on PW Doppler are calculated normally but not loaded correctly into hardware

Code information

All product lots of the Acuson NX3, model # 11235612 containing the following software version: VA10A, VA10B, VA10C, VA10D, VA10E, VA10F, VB20A, and VB20B;

Distribution pattern

Worldwide Distribution: US (nationwide) including states of: AL, AZ, AR, CA, CO, CT, DC, FL, GA, IL,IA, IN, KY, LA, MD, MA, MI, MN, MS, NE, NV, NH, NJ, NM, NY, NC, ND, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VT, VA, WA, WV, WI and WY; and countries of: Albania, Algeria, Angola, Argentina, Australia, Austria, Azerbaijan, Bahrain, Bangladesh, Belarus, Belgium, Bolivia, Bosnia, Brazil, Brunei, Bulgaria, Burkina Faso, Cameroon, Canada, Chile, China, Colombia, Costa Rica, Croatia, Cyrus, Czech Republic, Democratic Republic of Congo, Denmark, Dominican Republic, Ecuador, Egypt, El Salvador Finland, France, Gabon, Georgia, Germany, Greece, Guatemala, Hong Kong, Hungary, India, Indonesia, Iran, Iraq, Ireland, Israel, Italy, Ivory Coast, Japan, Jordan, Kazakhstan, Kenya, Kosovo, Latvia, Lebanon, Luxembourg, Macedonia, Malaysia, Maldives, Mali, Mexico, Morocco, Nepal, Netherlands, Norway, Pakistan, Palestine, Panama, Philippines, Poland, Portugal, Republic of Korea, Romania, Ruanda, Russian Federation, Saudi Arabia, Senegal, Serbia, Singapore, Slovakia, Slovenia, South Africa, Spain, Sri Lanka, Sweden, Switzerland, Taiwan, Tanzania, Thailand, Turkey, U.A.E., Uganda, Ukraine, United Kingdom, Uruguay, Uzbekistan and Vietnam.

device · product 4 of 4

ACUSON NX3 Elite Diagnostic Ultrasound System, Model # 11235411, RX Only, Siemens Medical Solutions, containing software versions: VA10A, VA10B, VA10C, VA10D, VA10E, VA10F, VB20A, and VB20B; UDI # 04056869004303 The ACUSON NX2 diagnostic ultrasound imaging systems are intended for the following applications: Cardiac, Fetal, Abdominal (including liver), Pediatric, Small Parts (Small Organ), Transcranial, OB/GYN, Pelvic, Urology, Vascular (including Peripheral Vessel) and Musculoskeletal applications. The systems also provide for the measurement of anatomical structures and for analysis packages that provide information that is used for clinical diagnosis purposes.

Z-2353-2019
Recall number
Z-2353-2019
Initiated
January 28, 2019
Classification
Class II
Status
Ongoing
Quantity
2262 systems

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
The action is being initiated due to internal testing which identified a possibility for transducers to exceed the acoustic output power (AOP) values defined . The transit voltage values based on PW Doppler are calculated normally but not loaded correctly into hardware

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Sourced

Software design

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

The action is being initiated due to internal testing which identified a possibility for transducers to exceed the acoustic output power (AOP) values defined . The transit voltage values based on PW Doppler are calculated normally but not loaded correctly into hardware

Code information

All product lots of the Acuson NX3, model # 11235411 containing the following software version: VA10A, VA10B, VA10C, VA10D, VA10E, VA10F, VB20A, and VB20B;

Distribution pattern

Worldwide Distribution: US (nationwide) including states of: AL, AZ, AR, CA, CO, CT, DC, FL, GA, IL,IA, IN, KY, LA, MD, MA, MI, MN, MS, NE, NV, NH, NJ, NM, NY, NC, ND, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VT, VA, WA, WV, WI and WY; and countries of: Albania, Algeria, Angola, Argentina, Australia, Austria, Azerbaijan, Bahrain, Bangladesh, Belarus, Belgium, Bolivia, Bosnia, Brazil, Brunei, Bulgaria, Burkina Faso, Cameroon, Canada, Chile, China, Colombia, Costa Rica, Croatia, Cyrus, Czech Republic, Democratic Republic of Congo, Denmark, Dominican Republic, Ecuador, Egypt, El Salvador Finland, France, Gabon, Georgia, Germany, Greece, Guatemala, Hong Kong, Hungary, India, Indonesia, Iran, Iraq, Ireland, Israel, Italy, Ivory Coast, Japan, Jordan, Kazakhstan, Kenya, Kosovo, Latvia, Lebanon, Luxembourg, Macedonia, Malaysia, Maldives, Mali, Mexico, Morocco, Nepal, Netherlands, Norway, Pakistan, Palestine, Panama, Philippines, Poland, Portugal, Republic of Korea, Romania, Ruanda, Russian Federation, Saudi Arabia, Senegal, Serbia, Singapore, Slovakia, Slovenia, South Africa, Spain, Sri Lanka, Sweden, Switzerland, Taiwan, Tanzania, Thailand, Turkey, U.A.E., Uganda, Ukraine, United Kingdom, Uruguay, Uzbekistan and Vietnam.

Field note

Send feedback