Skip to content
Recall Observatory FDA recall evidence

Exact FDA event identifier

Event 82940

1 recalled-product record grouped only because every row carries this exact official event ID.

Open FDA event record

Event summary

Timeline bucket
May 03, 2019
Product types
Device
Classifications
Class I
Statuses
Terminated
Recalling firm wording
Medline Industries Inc

Dossier provenance

Source snapshots represented here

  • openFDA Device Enforcement Report-date coverage June 20, 2012–July 08, 2026 Known gap: openFDA describes 2004-present coverage, but the current export has no report-date records in this interval.
  • openFDA Device Recall event_date_initiated coverage June 01, 1997–June 30, 2026

Enforcement coverage uses report dates; product initiation dates can precede those bounds. Device-enrichment coverage uses FDA event_date_initiated.

How coverage and grouping work

Complete imported group

Every recalled product in this event

1 official enforcement rows

Product dates, classifications, firms, and source wording remain attached to their individual rows. Opening this dossier never hides products that did not match a prior timeline filter.

device · product 1 of 1

Medline Primary Warmer Pack, Product No. DYNJ52510A. The Medline Primary Warmer Pack contains the recalled Vyaire enFlow Disposable Cartridge, which is the subject of this recall, and the recalled BD Alaris Pump Infusion Set (RES#82951) Product Usage: The Vyaire enFlow Disposable Cartridges contain a heating element that is used to help raise or maintain a patient's body temperature.

Z-2066-2019
Recall number
Z-2066-2019
Initiated
May 03, 2019
Classification
Class I
Status
Terminated
Recalling firm
Medline Industries Inc
Quantity
4,680 kits

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
The Vyaire enFlow disposable cartridge was recalled due to the potential for unsafe aluminum levels to elute into the infusate. The Use of the BD Alaris Pump Infusion Set may lead to serious adverse events for patients due to the incomplete occlusion caused by variation in tubing wall thickness.

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Sourced

Nonconforming Material/Component

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

The Vyaire enFlow disposable cartridge was recalled due to the potential for unsafe aluminum levels to elute into the infusate. The Use of the BD Alaris Pump Infusion Set may lead to serious adverse events for patients due to the incomplete occlusion caused by variation in tubing wall thickness.

Code information

Lots 2018090690 2018101190 2018101590 2018121190 2019013190 2019030490 2019040190

Distribution pattern

One distributor located in Illinois.

Field note

Send feedback

We'll only use this to respond to your feedback.