Skip to content
Recall Observatory FDA recall evidence

Exact FDA event identifier

Event 82878

1 recalled-product record grouped only because every row carries this exact official event ID.

Open FDA event record

Event summary

Timeline bucket
October 17, 2018
Product types
Device
Classifications
Class II
Statuses
Terminated
Recalling firm wording
Terumo Cardiovascular Systems Corporation

Dossier provenance

Source snapshots represented here

  • openFDA Device Enforcement Report-date coverage June 20, 2012–July 08, 2026 Known gap: openFDA describes 2004-present coverage, but the current export has no report-date records in this interval.
  • openFDA Device Recall event_date_initiated coverage June 01, 1997–June 30, 2026

Enforcement coverage uses report dates; product initiation dates can precede those bounds. Device-enrichment coverage uses FDA event_date_initiated.

How coverage and grouping work

Complete imported group

Every recalled product in this event

1 official enforcement rows

Product dates, classifications, firms, and source wording remain attached to their individual rows. Opening this dossier never hides products that did not match a prior timeline filter.

device · product 1 of 1

Terumo Advanced Perfusion System 1 Electronic Patient Gas System (EPGS), Electronic Gas Blender, Catalog #: 801188, UDI 00886799000588. Used in cardio-pulmonary bypass.

Z-1769-2019
Recall number
Z-1769-2019
Initiated
October 17, 2018
Classification
Class II
Status
Terminated
Quantity
72

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
An intermittent failure which causes a blender initialization fault upon start-up or during an EPGS calibration attempt leads to the Central Control Monitor control sliders being disabled, such that the gas system can only be controlled by the local control knobs.

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Sourced

Process control

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

An intermittent failure which causes a blender initialization fault upon start-up or during an EPGS calibration attempt leads to the Central Control Monitor control sliders being disabled, such that the gas system can only be controlled by the local control knobs.

Code information

01338 01339 01340 01341 01342 01349 01350 01351 01352 01353 01354 01355 01356 01357 01358 01360 01361 01362 01363 01365 01366 01367 01369 01370 01371 01372 01373 01374 01375 01376 01377 01378 01379 Lot/Serial 01380 01381 01382 01383 01384 01385 01386 01387 01388 01389 01394 01395 01396 01397 01398 01405 01406 01407 01408 01410 01411 01413 01414 01415 01416 01418 01420 01421 01422 01423 01424 01425 01426 Lot/Serial 01427 01428 01429 01430 01431 01432 01434 01435 1718 2317 4417 9999 UDI 00886799000588

Distribution pattern

Distribution US nationwide, including CA, CT, IL, IN, KY, MI, MN, MO, MT, NM, NY, OH, and WI. Foreign distribution to Canada, Mexico, and South Korea.

Field note

Send feedback

We'll only use this to respond to your feedback.