Event summary

Timeline bucket
January 02, 2019
Product types
Device
Classifications
Class III
Statuses
Terminated
Recalling firm wording
Becton, Dickinson and Company, BD Biosciences

Dossier provenance

Source snapshots represented here

  • openFDA Device Enforcement Report-date coverage June 20, 2012–July 08, 2026 Known gap: openFDA describes 2004-present coverage, but the current export has no report-date records in this interval.
  • openFDA Device Recall event_date_initiated coverage June 01, 1997–June 30, 2026

Enforcement coverage uses report dates; product initiation dates can precede those bounds. Device-enrichment coverage uses FDA event_date_initiated.

How coverage and grouping work

Complete imported group

Every recalled product in this event

1 official enforcement rows

Product dates, classifications, firms, and source wording remain attached to their individual rows. Opening this dossier never hides products that did not match a prior timeline filter.

device · product 1 of 1

BD FACSLyric 3L 10C with FACSuite Clinical Software v1.1.1, catalog number 662878 BD FACSLyric is a high-performance flow cytometer designed to support both routine clinical analysis and research for the identification, quantification, and characterization of cells in support of cell analysis. BD FACSuite software is loaded onto an adjacent PC workstation and is used to operate the instrument, acquire samples, and analyze data. Quality control performance, tracking, and reporting are streamlined and automated. Routine tasks such as daily cleaning and shutdown can be programmed to occur automatically. The BD FACSuite clinical software allows users to run BD IVD assays.

Z-0792-2019
Recall number
Z-0792-2019
Initiated
January 02, 2019
Classification
Class III
Status
Terminated
Quantity
3 devices

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
A customer site was identified to have three FACSLyric IVD systems with FACSuite clinical v1.1.1 installed on a computer workstation with windows 10 operating system (OS). Windows 10 OS is not a released configuration for clinical use. It is released for research use only.

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Unknown

An exact joined enrichment record exists, but none supplies supported root-cause wording. This is not an FDA finding of an unknown cause.

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

A customer site was identified to have three FACSLyric IVD systems with FACSuite clinical v1.1.1 installed on a computer workstation with windows 10 operating system (OS). Windows 10 OS is not a released configuration for clinical use. It is released for research use only.

Code information

UDI: (01)00382906628787(11)180508(21)Z662878000056 serial number: Z662878000056 Work Station SN 2UA80820ZZ UDI: (01)00382906628787(11)180301(21)Z662878000029 serial number: Z662878000029 Work Station SN 2UA80820ZW UDI: (01)00382903389605(11)180315(21)11662382000007 serial number: R662382000007 Work Station SN 2UA8082101

Distribution pattern

US Distribution in MD.

Field note

Send feedback