device · product 1 of 1
UroLift System REF Catalog Number: UL400-4 Tray Contents: UroLift System with Handle Release Tool STERILE R CE 0050 Manufactured By: neotract
- Recall number
- Z-0598-2019
- Initiated
- July 06, 2018
- Classification
- Class II
- Status
- Terminated
- Recalling firm
- Neotract Inc
- Quantity
- 8,804
App-derived interpretation
Upon implant deployment, the Capsular Tab may not be delivered as the needle is retracted. In this failure mode, the needle is deployed into the prostate and retracted, leaving no implant behind. Failure to deliver a capsular tab may result in a delay in completing a treatment or an inability to complete a treatment for the patient. Use of a recalled device may lead to increased frequency of exiting known risks associated with the product, including bleeding and tissue trauma associated with delivering the needle.
Official device-enrichment evidence · Sourced
Finished device change control
Inspect official wording and provenance
Reason for recall
Upon implant deployment, the Capsular Tab may not be delivered as the needle is retracted. In this failure mode, the needle is deployed into the prostate and retracted, leaving no implant behind. Failure to deliver a capsular tab may result in a delay in completing a treatment or an inability to complete a treatment for the patient. Use of a recalled device may lead to increased frequency of exiting known risks associated with the product, including bleeding and tissue trauma associated with delivering the needle.
Code information
Model: AN00141 (US) and AN00142 (OUS); Catalog No.: UL400-4; Lots: P37940, P37943, P37953, P37956, P37965, P37971, P37985, P37988, P37995, P38001, P38013, P38020, P38124, P38132
Distribution pattern
US: AK, AL, AZ, CA, CO, CT, DE, FL, GA, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, TN, TX, UT, VA, WA, WI, WV,