Event 81419

1 recalled-product record grouped only because every row carries this exact official event ID.

Open FDA event record

Event summary

Timeline bucket
October 05, 2018
Product types
Device
Classifications
Class II
Statuses
Terminated
Recalling firm wording
Ohmeda Medical

Dossier provenance

Source snapshots represented here

  • openFDA Device Enforcement Report-date coverage June 20, 2012–July 08, 2026 Known gap: openFDA describes 2004-present coverage, but the current export has no report-date records in this interval.
  • openFDA Device Recall event_date_initiated coverage June 01, 1997–June 30, 2026

Enforcement coverage uses report dates; product initiation dates can precede those bounds. Device-enrichment coverage uses FDA event_date_initiated.

How coverage and grouping work

Complete imported group

Every recalled product in this event

1 official enforcement rows

Product dates, classifications, firms, and source wording remain attached to their individual rows. Opening this dossier never hides products that did not match a prior timeline filter.

device · product 1 of 1

Giraffe Omnibed - Product Usage: The OmniBed is a combination of an infant incubator and an infant warmer. The device can be operated as an incubator or as a warmer and can transition from one mode to the other on user s demand. It cannot be operated in both modes at the same time. Incubators and warmers provide heat in a controlled manner to neonates who are unable to thermo-regulate based on their own physiology. Incubators provide an enclosed, temperature-controlled environment and warmers provide infrared heat in an open environment. They may also be used for short periods of time to facilitate the neonate s transition from the uterus to the external environment. This device may incorporate a Servo Controlled Oxygen Delivery System. This is indicated to provide stable oxygen concentration within the infant compartment at the value set by the operator (21-65%).

Z-0041-2020
Recall number
Z-0041-2020
Initiated
October 05, 2018
Classification
Class II
Status
Terminated
Recalling firm
Ohmeda Medical
Quantity
N/A

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
GE Healthcare has become aware of a potential safety issue that can occur if the canopy seals are not cleaned appropriately. Healthcare professionals reported buildup of debris around the edge of the canopy seal which can lead to a source of infection. Upon review, the Agency concluded that the cleaning and care guidelines provided by the firm were inadequate for effective cleaning of the device. In particular, the instructions do not instruct for removal of canopy seal prior to cleaning, which leads to the buildup of debris.

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Sourced

Labeling design

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

GE Healthcare has become aware of a potential safety issue that can occur if the canopy seals are not cleaned appropriately. Healthcare professionals reported buildup of debris around the edge of the canopy seal which can lead to a source of infection. Upon review, the Agency concluded that the cleaning and care guidelines provided by the firm were inadequate for effective cleaning of the device. In particular, the instructions do not instruct for removal of canopy seal prior to cleaning, which leads to the buildup of debris.

Code information

Giraffe OmniBeds shipped from 2000 to 2011 with Serial Numbers starting from HDGD00001 to HDGQ50183.

Distribution pattern

Worldwide distribution- Domestic nationwide and foreign distribution.