Skip to content
Recall Observatory FDA recall evidence

Exact FDA event identifier

Event 81169

1 recalled-product record grouped only because every row carries this exact official event ID.

Open FDA event record

Event summary

Timeline bucket
October 17, 2018
Product types
Device
Classifications
Class II
Statuses
Ongoing
Recalling firm wording
Richard Wolf GmbH

Dossier provenance

Source snapshots represented here

  • openFDA Device Enforcement Report-date coverage June 20, 2012–July 08, 2026 Known gap: openFDA describes 2004-present coverage, but the current export has no report-date records in this interval.
  • openFDA Device Recall event_date_initiated coverage June 01, 1997–June 30, 2026

Enforcement coverage uses report dates; product initiation dates can precede those bounds. Device-enrichment coverage uses FDA event_date_initiated.

How coverage and grouping work

Complete imported group

Every recalled product in this event

1 official enforcement rows

Product dates, classifications, firms, and source wording remain attached to their individual rows. Opening this dossier never hides products that did not match a prior timeline filter.

device · product 1 of 1

Richard Wolf Bronchoscope Tubes, Model Texas. Device is reusable and should be sterilized prior to each use. Product Usage: Bronchoscopes are intended for investigation, diagnosis and therapy of pathological changes and removal of foreign bodies in the area of the tracheobronchial respiratory pathways, in conjunction with endoscopic accessories. The bronchoscopes can also be used endoscopically to visualize the tracheobionchial respiratory tract via the natural passage (oral). Tracheoscope tubes are used as follows: 1. The tube is used as a passage to the area to be examined or treated and allows the insertion of auxiliary instruments such as forceps, resector, remover , suction device and stents; 2. The internal diameter allows the conveying of gases for artificial respiration of the patient and via the ventilation connector in conjunction with the telescope element, the tube can be connected to all known respiration and anasthetic systems; 3. Sampling channel for respiratory gas analysis when gas for analysis is gained from the patient s central respiratory tract by means of the bypass method via the tracheoscope tube. M = identification on Luer connector; 4.Tracheoscopes are used in the trachea up to the tracheal bifurcation.

Z-0804-2019
Recall number
Z-0804-2019
Initiated
October 17, 2018
Classification
Class II
Status
Ongoing
Recalling firm
Richard Wolf GmbH
Quantity
873 units worldwide (468 units within the US)

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
There is a potential deficiency of the TEXAS Bronchoscope Tube product. The non-functionally relevant bonding points / cemented areas of the Bronchoscope Tubes may loosen.

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Sourced

Process control

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

There is a potential deficiency of the TEXAS Bronchoscope Tube product. The non-functionally relevant bonding points / cemented areas of the Bronchoscope Tubes may loosen.

Code information

All lot numbers manufactured from October 30, 2006 to August 15, 2018, with affected Product and Lot Numbers distributed worldwide - Product Number (Lot Numbers): 825200641 (1271924, 1074548, 1186516, 1364805, 1262505, 1269165, 1313148, 1296901, 1330427, 1382941, 1321674, 1339578, 1265501, 1344236, 1261415, 1372437, 1353961, M054680, M118520, 1003492, 1058979, 1287750); 825200842 (1189516, 1074551, 1319855, 1298913, 1057370, M054700, M118540); 825201041 (1269824, 1277860, 1186520, 1195261, 1230705, 1340401, 1253661, 1263078, 1288520, 1186520 , 1227431, 1368550, 1376018, 1382289, 1316167, 1104163, 1074553, 1307329, 1296285, 1332473, 1319856, 1053400, 1337541, M054720, M118550); 825201241 (1185831, 1372500, 1364881, 1238365, 1273017, 1302457, 1298771, 1186521, 1269168, 1183846, 1323532, 1074555, 1316169, 1344797, 1333774, 1057371, 1361470, 1265502, 1289744, M054730, M118560); 825201441 (1185832, 1189518, 1339287, 1274102, 1281289, 1320136, 1314041, 1360269, 1328548, 1003493, 1074556, M054740, M112910); 825210830 (1264551, 1313443, 1338468, 1212278, 1262656, 1271357, 1320727); 825211030 (1271954, 1293167, 1287813, 1313444, 1265025, 1320728, 1212279, 1342362); 825211230 (1240940, 1253421, 1245260); 825211430 (1253422, 1240941); 825200841 (1274105, 1185829, 1281667, 1257138, 1368638, 1306948, 1265989, 1296284, 1298436, 1189515, 1269820, 1319630, 1376097, 1362251, 1323968, 1107980, 1333771, 1261416, 1378166, 1053398, 1338467, 1288856, 1074550, M054690, M118530, 1241165, 1243611, 1248730, 1253658, 1262506); Affected Part and Lot numbers distributed Nationwide - Part Number (Lot Numbers): 82520.0641 (1058979, 1074548, 1339578, 1003492, 1313148, M054680, 1186516, 1271924, 1262505, 1269165, 1296901, 1364805, 1330427, 1344236, M118520, 1287750, 1353961, 1382941); 82520.0841 (1074550, 1333771, 1323968, M054690, 1281667, 1274105, 1241165, 1053398, 1269820, 1243611, 1248730, 1262506, 1288856, 1298436, M118530, 1362251, 1185829, 1198515, 1338467, 1378166, 1253658, 1296284); 82520.0842 (M118540, M054700, 1189516, 1319855); 82520.1041 (1074553, 1319856, 1316167, 1277860, M054720, 1053400, 1269824, 1230705, 1253661, 1263078, 1376018, 1288520, 1332473, M118550, 1337541, 1186520, 1195261, 1382289); 82520.1241 (1074555, 1333774, 1302457, 1057371, M054730, 1269168, M118560, 1238365, 1186521, 1289744, 1372500, 1298771, 1344797, 1361470, 1265502, 1185831); 82520.1441 (1074556, 1320136, 1003493, 1314041, 1281289, M054740, M112910, 1189518, 1360269, 1274102, 1339287, 1328548); 825210830 (1271357, 1313443, 1212278, 1262656, 1320727, 1264551, 1338468); 825211030 (1320728, 1313444, 1212279, 1265025, 1342362, 1293167, 1287813); 825211230 (1245260, 1253421, 1240940); 825211430 (1240941, 1253422)

Distribution pattern

Worldwide distribution - US Nationwide in the states of PA, MD, NY, and countries of Bangladesh, Canada, China, France, Germany, India, Italy, Jordan, Netherlands, Poland, Russia, Spain, Thailand, Turkey.

Field note

Send feedback

We'll only use this to respond to your feedback.