openFDA Device Enforcement
Report-date coverage
June 20, 2012–July 08, 2026
Known gap: openFDA describes 2004-present coverage, but the current export has no report-date records in this interval.
openFDA Device Recall
event_date_initiated coverage
June 01, 1997–June 30, 2026
Enforcement coverage uses report dates; product initiation dates can precede those bounds. Device-enrichment coverage uses FDA event_date_initiated.
Product dates, classifications, firms, and source wording remain attached to their individual rows.
Opening this dossier never hides products that did not match a prior timeline filter.
device · product 1 of 1
Sysmex XN-Series Fluorocell PLT, Catalog #CD994563. Distributed in the U.S.A. by Sysmex America, Inc., Lincolnshire, IL. Fluorocell PLT is used to stain platelets in a diluted blood sample for the counting of platelets with the Sysmex XN-Series automated hematology analyzers.
The fluorescent platelet counts (PLT-F) and immature platelet fraction (IPF) values can be falsely low when using affected lots of Fluorocell PLT.
Code information
Reagent lot numbers: A7125, exp. date 01/9/2019; A7126, exp. date 01/11/2019; A7127, exp. date 01/09/2019, A7128, exp. 01/15/2019; A7129, exp. 01/15/2019; A7130, exp. 01/17/2019; A7131, exp. 01/17/2019, A8003, exp. 01/31/2019; A8004, exp. 01/31/2019; A8005, exp. 02/04/2019; and A8006, exp. 02/04/2019.
Distribution pattern
Worldwide Distribution: US (Nationwide) including Puerto Rico; and countries of: Argentina, Brazil, Canada, Chile, Costa Rica, Ecuador, Honduras, and Nicaragua.