Event 80512

1 recalled-product record grouped only because every row carries this exact official event ID.

Open FDA event record

Event summary

Timeline bucket
June 11, 2018
Product types
Device
Classifications
Class II
Statuses
Terminated
Recalling firm wording
NxStage Medical, Inc.

Dossier provenance

Source snapshots represented here

  • openFDA Device Enforcement Report-date coverage June 20, 2012–July 08, 2026 Known gap: openFDA describes 2004-present coverage, but the current export has no report-date records in this interval.
  • openFDA Device Recall event_date_initiated coverage June 01, 1997–June 30, 2026

Enforcement coverage uses report dates; product initiation dates can precede those bounds. Device-enrichment coverage uses FDA event_date_initiated.

How coverage and grouping work

Complete imported group

Every recalled product in this event

1 official enforcement rows

Product dates, classifications, firms, and source wording remain attached to their individual rows. Opening this dossier never hides products that did not match a prior timeline filter.

device · product 1 of 1

NxStage Express Fluid Warmer (FW-300, FW-300-A, FW-300-WW, FW-301, FW-301-A) Product Usage: The NxStage Express Fluid Warmer is an accessory to the NxStage System One used to warm therapy fluids prior to administration.

Z-2793-2018
Recall number
Z-2793-2018
Initiated
June 11, 2018
Classification
Class II
Status
Terminated
Recalling firm
NxStage Medical, Inc.
Quantity
14975

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
There is a potential risk of electrical fire when fluid leaks into the power receptacle on the back of the warmer. The risk exists any time the warmer is plugged in. The firm has received 7 reports of smoke, sparks, or flames coming from the Express Fluid Warmer since 2016. No adverse events have been reported.

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Sourced

Use error

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

There is a potential risk of electrical fire when fluid leaks into the power receptacle on the back of the warmer. The risk exists any time the warmer is plugged in. The firm has received 7 reports of smoke, sparks, or flames coming from the Express Fluid Warmer since 2016. No adverse events have been reported.

Code information

All units are affected. As of 6/11/18, the last serial number manufactured was W35452.

Distribution pattern

Worldwide Distribution - US Nationwide in the countries of Australia, Canada, France, Germany, Italy, Kuwait, Netherlands, Saudi Arabia, Spain, Sweden, and UK.