device · product 1 of 1
PR4100 Microplate Reader US, Catalog Number 45591587 Product Usage: The PR4100 Microplate Reader is a 96 well absorbance reader for the measurement of light absorbance (optical density-OD) of liquid media. The instrument is intended to be used primarily for the in vitro diagnostic analysis of samples from the human body to obtain information oh physiological and pathological states.
- Recall number
- Z-2541-2018
- Initiated
- April 02, 2018
- Classification
- Class II
- Status
- Terminated
- Recalling firm
- Bio-Rad Laboratories Inc
- Quantity
- 153
App-derived interpretation
The connector of the power supply was capable of fitting other components provided with the system. However, the voltage output of the reader power supply is incompatible with the other component and, if incorrectly connected, could result in irreparable damage to the system.
Official device-enrichment evidence · Sourced
Device Design
Inspect official wording and provenance
Reason for recall
The connector of the power supply was capable of fitting other components provided with the system. However, the voltage output of the reader power supply is incompatible with the other component and, if incorrectly connected, could result in irreparable damage to the system.
Code information
UDI: 00847865014151 Serial Numbers: 140600072, 160900991, 161001953, 1112001938, 1112001941 , 1112004075 , 1112004079 , 1112004083 , 1112004084 , 1112004086 , 1112004088 , 1112004093 , 1201003822 , 1201003826 , 1201003827 , 1207001007 , 1207001015 , 1207001017 , 1207001018 , 1208001583 , 1208001584 , 1208001586 , 1208001590 , 1208005236 , 1208005244 , 1208005245 , 1211000978 , 1211000983 , 1211006938 , 1301003205 , 1301003212 , 1303000299 , 1303000300 , 1303000302 , 1303000305 , 1303007688 , 1306003195 , 1307006044 , 1307006045 , 1307006054 , 1307006055 , 1308000967 , 1308000968 , 1308000969 , 1308000971 , 1308000973 , 1308003338 , 1308003341 , 1308003345 , 1308003346 , 1308006670 , 1311008542 , 1311008544 , 1311008546 , 1311008555 , 1311008557 , 1311008562 , 1311008564 , 1311008565 , 1401002580 , 1401002583 , 1401002587 , 1401007785 , 1401007786 , 1401007789 , 1401007792 , 1401007793 , 1404000065 , 1404000066 , 1404000067 , 1404000068 , 1404002677 , 1404002684 , 1404002685 , 1405005640 , 1405005641 , 1405005645 , 1405005648 , 1406000064 , 1406000065 , 1406000066 , 1406000067 , 1406000068 , 1406000070 , 1406000071 , 1406000074 , 1406006883 , 1406006886 , 1407004144 , 1407004145 , 1407004147 , 1407004149 , 1407004151 , 1407004152 , 1410010410 , 1410010415 , 1410010416 , 1411010876 , 1411010877 , 1411010880 , 1412001714 , 1412001716 , 1412001721 , 1501001523 , 1501001526 , 1501001526 , 1501005823 , 1501005824 , 1501005825 , 1501005828 , 1501005829 , 1501005832 , 1506008687 , 1506008690 , 1506008695 , 1510005389 , 1510005393 , 1510005394 , 1510005395 , 1510005397 , 1510005398 , 1510005399 , 1510005414 , 1601001552 , 1601001553 , 1601001559 , 1601001561 , 1601001562 , 1601001563 , 1603001786 , 1603005407 , 1605000088 , 1609000992 , 1609000993 , 1609000994 , 1609000995 , 1609000996 , 1609001301 , 1609001302 , 1609001303 , 1609001304 , 1609001305 , 1609001306 , 1609009231 , 1610001955 , 1701001785 , 1701001788 , 1701004204 , 1701004210 , 1710010533 , 1710010539 , 13060003204 , 16303005411 , s1411010876
Distribution pattern
US Nationwide Distribution inn the states of : AL, AR, AZ, CA, CO, CT, FL, GA, HI, IL, IN, KS, LA, MA, MD, MI, MN, MO, MS, NC, NJ, NY, OR, PA, SC, TN, TX, VA, VT, WA, WI, and Puerto Rico.