Skip to content
Recall Observatory FDA recall evidence

Exact FDA event identifier

Event 80159

2 recalled-product records grouped only because every row carries this exact official event ID.

Open FDA event record

Event summary

Timeline bucket
January 25, 2018
Product types
Device
Classifications
Class II
Statuses
Terminated
Recalling firm wording
GE Healthcare

Dossier provenance

Source snapshots represented here

  • openFDA Device Enforcement Report-date coverage June 20, 2012–July 08, 2026 Known gap: openFDA describes 2004-present coverage, but the current export has no report-date records in this interval.
  • openFDA Device Recall event_date_initiated coverage June 01, 1997–June 30, 2026

Enforcement coverage uses report dates; product initiation dates can precede those bounds. Device-enrichment coverage uses FDA event_date_initiated.

How coverage and grouping work

Complete imported group

Every recalled product in this event

2 official enforcement rows

Product dates, classifications, firms, and source wording remain attached to their individual rows. Opening this dossier never hides products that did not match a prior timeline filter.

device · product 1 of 2

Centricity PACS-IW with Universal Viewer Product Usage: Usage: Centricity PACS-IW with Universal Viewer is a device that displays medical images (including mammograms) and data from various imaging sources. Images and data can be viewed, communicated, processed and displayed within the system or across computer networks at distributed locations. Lossy compressed mammographic images and digitized film screen images must not be reviewed for primary image interpretations. Mammographic images may only be interpreted using an FDA approved monitor that offers at least 5 Megapixel resolution and meet other technical specifications reviewed and accepted by the FDA. Typical users of this system are trained professionals, including but not limited to radiologists, physicians, nurses, medical technicians, and assistants.

Z-2101-2018
Recall number
Z-2101-2018
Initiated
January 25, 2018
Classification
Class II
Status
Terminated
Recalling firm
GE Healthcare
Quantity
44

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
Potential that one or more image series may be missing from an exam without a user warning displayed in the Viewer.

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Sourced

Software design

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

Potential that one or more image series may be missing from an exam without a user warning displayed in the Viewer.

Code information

102747-1-Centricity Universal Viewer-00558673, 11580-1-Centricity Universal Viewer-00558646, 12945-1-Centricity Universal Viewer-00558658, 103852-1-Centricity Universal Viewer-00116028, 103246-1-Centricity Universal Viewer-00558668, 100707-1-Centricity Universal Viewer-00558649, 1045-1-Centricity Universal Viewer-00558651, 101725-2-Centricity Universal Viewer-00558642, 10227-1-Centricity Universal Viewer-00558665, 9768-1-Centricity Universal Viewer-00558661, 100019-1-Centricity Universal Viewer-00558666, 4284-1-Centricity Universal Viewer-00558672, 101272-1-Centricity Universal Viewer-00558671, 100862-1-Centricity Universal Viewer-01444483, 10278-1-Centricity Universal Viewer-00558655, 9789-1-Centricity Universal Viewer-00558645, 12914-1-Centricity Universal Viewer-00558660, 1696-1-Centricity Universal Viewer-00558659, 4408-1-Centricity Universal Viewer-00558664, 4410-1-Centricity Universal Viewer-00558656, 4407-1-Centricity Universal Viewer-00558657, 4407-1-Centricity Universal Viewer-00558652, 4405-1-Centricity Universal Viewer-00558662, 4546-1-Centricity Universal Viewer-00558670, 100064-1-Centricity Universal Viewer-00558572, 100092-1-Centricity Universal ,Viewer-00558648, 1345-1-Centricity Universal Viewer-00558654, 4705-1-Centricity Universal Viewer-00558663, 4418-1-Centricity Universal Viewer-00558674, 102062-1-Centricity Universal Viewer-00558650, 12866-1-Centricity Universal Viewer-01448215, 101931-1-Centricity Universal Viewer-00558569, 4601-1-Centricity Universal Viewer-00558710, 102406-2-Centricity Universal Viewer-00558669, 102406-2-Centricity Universal Viewer-00558667 A59927-Centricity Universal Viewer-00040382, X13479-Centricity Universal Viewer-00043936, and 269338-Centricity Universal Viewer-01453050, 94767-Centricity Universal Viewer-01422997.

Distribution pattern

Worldwide Distribution - US Nationwide in the states of AK, AZ, CA, CO, DE, FL, GA, HI, IA, ID, IL, LA, MA, MN, MO, MS, MT, NC, NE, NH, NJ, NM, NV, NY, OK, OR. PA. SD, TN, TX, UT, WA, WI, WV, and WY and the countries of Austria, Belgium, Brazil, Canada, Colombia, France, Germany, Ghana, India, Ireland, Israel, Italy, Kuwait, Malaysia, Mexico, Netherlands, Panama, Poland, Russia, Saudi Arabia, South Africa, Spain, Switzerland, Turkey, United Arab Emirates, United Kingdomand Uruguay

device · product 2 of 2

Centricity Universal Viewer Product Usage: Usage: Centricity PACS-IW with Universal Viewer is a device that displays medical images (including mammograms) and data from various imaging sources. Images and data can be viewed, communicated, processed and displayed within the system or across computer networks at distributed locations. Lossy compressed mammographic images and digitized film screen images must not be reviewed for primary image interpretations. Mammographic images may only be interpreted using an FDA approved monitor that offers at least 5 Megapixel resolution and meet other technical specifications reviewed and accepted by the FDA. Typical users of this system are trained professionals, including but not limited to radiologists, physicians, nurses, medical technicians, and assistants.

Z-2102-2018
Recall number
Z-2102-2018
Initiated
January 25, 2018
Classification
Class II
Status
Terminated
Recalling firm
GE Healthcare
Quantity
252

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
Potential that one or more image series may be missing from an exam without a user warning displayed in the Viewer.

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Sourced

Software design

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

Potential that one or more image series may be missing from an exam without a user warning displayed in the Viewer.

Code information

102123-1-Centricity Universal Viewer-00558570, 9796-1-Centricity Universal Viewer-00558571, 1011-1-Centricity Universal Viewer-00558109, 4433-1-Centricity Universal Viewer-00558704, 4446-2-Centricity Universal Viewer-00558643, 13240-2-Centricity Universal Viewer-00558103, 4487-1-Centricity Universal Viewer-01291589, 13295-1-Centricity Universal Viewer-01292684, 103090-2-Centricity Universal Viewer-00546048, 3760-1-Centricity Universal Viewer-00101627, 102580-1-Centricity Universal Viewer-00558689, 102143-1-Centricity Universal Viewer-00558685, 4271-1-Centricity Universal Viewer-00558351, 4273-3-Centricity Universal Viewer-00558104, 2440-1-Centricity Universal Viewer-00558701, 100366-1-Centricity Universal Viewer-00558522, 100769-1-Centricity Universal Viewer-01298556, 102278-1-Centricity Universal Viewer-00558682, 3846-3-Centricity Universal Viewer-00111167, 4449-1-Centricity Universal Viewer-00558276, 3245-1-Centricity Universal Viewer-01451973, 101690-1-Centricity Universal Viewer-00558680, 101039-1-Centricity Universal Viewer-01373579, 4451-1-Centricity Universal Viewer-,00558107, 100837-1-Centricity Universal Viewer-00558435, 4583-1-Centricity Universal Viewer-01367198, 13162-8-Centricity Universal Viewer-01452133, 4491-1-Centricity Universal Viewer-00558690, 13296-1-Centricity Universal Viewer-01364047, 9752-1-Centricity Universal Viewer-00558454, 1472-15-Centricity Universal Viewer-01367676, 101044-1-Centricity Universal Viewer-00558694, 13228-1-Centricity Universal Viewer-00558678, 102335-1-Centricity Universal Viewer-00558675, 4369-1-Centricity Universal Viewer-00558705, 4615-1-Centricity Universal Viewer-01452088, 4371-1-Centricity Universal Viewer-00558358, 4289-1-Centricity Universal Viewer-00558459, 101274-1-Centricity Universal Viewer-00558679, 100693-1-Centricity Universal Viewer-00558684, 104491-1-Centricity Universal Viewer-01288014, 1027-1-Centricity Universal Viewer-00558706, 104336-1-Centricity Universal Viewer-00346858, 4578-1-Centricity Universal Viewer-00558340, 13345-1-Centricity Universal Viewer-00602707, 4375-1-Centricity Universal Viewer-00558392, 9852-2-Centricity Universal Viewer-00558644, 101726-1-Centricity Universal Viewer-00558688, 100256-1-Centricity Universal Viewer-00558700, 4308-1-Centricity Universal Viewer-00558676, 100773-1-Centricity Universal Viewer-00558484, 4669-1-Centricity Universal Viewer-00558691, 100159-1-Centricity Universal Viewer-00558324, 4406-1-Centricity Universal Viewer-00558683, 4314-1-Centricity Universal Viewer-00558466, 4318-1-Centricity Universal Viewer-00558703, 102325-1-Centricity Universal Viewer-00558647, 4415-1-Centricity Universal Viewer-00558359, 4585-3-Centricity Universal Viewer-01448365, 4440-1-Centricity Universal Viewer-00558353, 4441-1-Centricity Universal Viewer-00558677, 104558-1-Centricity Universal Viewer-01522793, 4693-1-Centricity Universal Viewer-00558687, 104120-1-Centricity Universal Viewer-00164159, 100717-1-Centricity Universal Viewer-01293807, 4048-1-Centricity Universal Viewer-00558695, 102437-1-Centricity Universal Viewer-00558698, 4698-1-Centricity Universal Viewer-00558313, 13356-2-Centricity Universal Viewer-00558681, 12902-1-Centricity Universal Viewer-00602638, 100614-1-Centricity Universal Viewer-00558693, 4325-51-Centricity Universal Viewer-00599129, 101120-1-Centricity Universal Viewer-00558702, 4483-1-Centricity Universal Viewer-00558457, 102406-2-Centricity Universal Viewer-00558111, 102406-2-Centricity Universal Viewer-00558692, 102406-2-Centricity Universal Viewer-00558686, 102406-2-Centricity Universal Viewer-00558707, 13125-1-Centricity Universal Viewer-00558110, 102693-1-Centricity Universal Viewer-01282651, 101598-1-Centricity Universal Viewer-00667138, 318055-Centricity Universal Viewer-01457316, 85021-Centricity Universal Viewer-01426096, AE1476-Centricity Universal Viewer-01509883, AE1478-Centricity Universal Viewer-00163417, 06051-Centricity Universal Viewer-00115203, 48010-Centricity Universal Viewer-01444602, 34381-Centricity Universal Viewer-00116124, 53008-Centricity Universal Viewer-00116118, 482896-Centricity Universal Viewer-00545574, 228493-Centricity Universal Viewer-00599935, 802030-Centricity Universal Viewer-00000462, 899010-Centricity Universal Viewer-01300907, 123953-Centricity Universal Viewer-00163388, 651948-Centricity Universal Viewer-01813739, YP3150-Centricity Universal Viewer-01639018, FP5028-Centricity Universal Viewer-00110911, ZA00113-Centricity Universal Viewer-01474463, ZA2533-Centricity Universal Viewer-00116577, ZA1058-Centricity Universal Viewer-01724673, ZA2551-Centricity Universal Viewer-00116834, SA1339-Universal Viewer-00116444, RU7175-Centricity Universal Viewer-01445748, RU8414-Centricity Universal Viewer-01445730, RU7808-Centricity Universal Viewer-01445764, RU1745-Centricity Universal Viewer-01445766, RU3998-Centricity Universal Viewer-01553790, SiteID-Centricity Universal Viewer-01509505, RU8002-Centricity PACS-IW-01273292, RU1646-Centricity Universal Viewer-01445763, RU2580-Centricity Universal Viewer-01445743, SiteID-Centricity Universal Viewer-01476628, 600101-Centricity Universal Viewer-01273190, RU8356-Centricity Universal Viewer-01445768, RU9059-Centricity Universal Viewer-01445746, RU1112-Centricity Universal Viewer-01453917, RU3866-Centricity Universal Viewer-01445731, RU3453-Centricity Universal Viewer-01273249, RU1011-Centricity Universal Viewer-01445747, RU1130-Centricity Universal Viewer-01273228, Centricity Universal Viewer-01462487, RU5388-Centricity Universal Viewer-01719556, RU1106-Centricity Universal Viewer-01445765, RU8945-Centricity Universal Viewer-01445740,PL3249-Centricity Universal Viewer-00038866, PL3673-Centricity Universal Viewer-00038810, PL1494-Centricity Universal Viewer-00038799, PL2970-Centricity Universal Viewer-00038797, PL1499-Centricity Universal Viewer-00038800, PL1501-Centricity Universal Viewer-00038802, PL1124-Centricity Universal Viewer-00038795, PL2624-Centricity Universal Viewer-00038804, PL1504-Centricity Universal Viewer-00038806, PL2971-Centricity Universal Viewer-00038808, PL3277-Centricity Universal Viewer-00038831, PL2975-Centricity Universal Viewer-00162605 PL3349-Centricity Universal Viewer-00038811, AMS39-Centricity Universal Viewer-00545606, BON01-Centricity Universal Viewer-00044131, ECT02-Centricity Universal Viewer-00059842, 338896-Centricity PACS-IW-01443491, 329052-Centricity Universal Viewer-01443407, HUSM-Centricity Universal Viewer-01286208, IPPT-Centricity Universal Viewer-01286213, KW1005-Centricity Universal Viewer-01300924, KW0000-Centricity Universal Viewer-01416913, A51911-Centricity Universal Viewer-00040346, A54120-Centricity Universal Viewer-00040449, A3790481-Centricity Universal Viewer-00040932 A3784447-Centricity Universal Viewer-00158569, IL1469-Centricity Universal Viewer-00040970, 81111-Centricity Universal Viewer-01286673, 0000220630II8, 00000230818II7, SiteID-Centricity Universal Viewer-01724675, 434487-Centricity Universal Viewer-01453839, UC2011-Centricity Universal Viewer-01282271, 786787-Centricity Universal Viewer-01283171, PC1961-C-Generic Web Client-00001718, PC1961-Centricity Universal Viewer-00164218, PC1961-Centricity Universal Viewer-00001720, HBON01-Centricity Universal Viewer-00260801, HC9502-Centricity Universal Viewer-00260671, PC1695-Centricity Universal Viewer-01451951, PC1695-Centricity Universal Viewer-01447701, HC0245-Centricity Universal Viewer-00059319, 691749-Centricity Universal Viewer-01539092, HC1642-Centricity Universal Viewer-01634899, HC3062-Centricity Universal Viewer-01462397, HC4105-Centricity Universal Viewer-00001393, CC4105-Centricity Universal Viewer-01282278, 233637-Centricity Universal Viewer-01444341, 485135-Centricity Universal Viewer-01447706, HC0565-Centricity Universal Viewer-00545595, 530366-Centricity Universal Viewer-00260659, PC0023-Centricity Universal Viewer-01282279, PX4900-Centricity Universal Viewer-01369835, 601860-Centricity Universal Viewer-01373156, HC0529-Centricity Universal Viewer-01364776, 531344-Centricity Universal Viewer-00000707, HG1900-Centricity Universal Viewer-01282276, CC0154-Centricity Universal Viewer-01813738, UHAM02-Centricity Universal Viewer-01447361, UHAM02-Centricity Universal Viewer-01447361, M41613-Centricity Universal Viewer-00043666, M44610-Centricity Universal Viewer-00043788, E41676-Centricity Universal Viewer-00043587, X3141198-Centricity Universal Viewer-00043959, M15393-Centricity Universal Viewer-00043611, M41745-Centricity Universal Viewer-00043708, X1413856-Centricity Universal Viewer-00044474, X1982342-Centricity Universal Viewer-00043950, M45006-Centricity Universal Viewer-00043854, M44917-Centricity Universal Viewer-00043845, M46163-Centricity Universal Viewer-00043858, M41765-Centricity Universal Viewer-00043715, A54852-Centricity Universal Viewer-00043560, M41870-Centricity Universal Viewer-00043721, M41894-Centricity Universal Viewer-00043726, M42073-Centricity Universal Viewer-00044530, M60364-Centricity Universal Viewer-01801169, M44875-Centricity Universal Viewer-00043829, M41913-Centricity Universal Viewer-00065993, M41914-Centricity Universal Viewer-00043730, X3041774-Centricity Universal Viewer-00043956, X2015148-Centricity Universal Viewer-00044553, M44599-Centricity Universal Viewer-00043784, M232471-Centricity Universal Viewer-00043629, A91901-Centricity Universal Viewer-00044549, M41980-Centricity Universal Viewer-00044515, M41981-Centricity Universal Viewer-01270561, M41974-Centricity Universal Viewer-00043736, M43605-Centricity Universal Viewer-00043769, M44877-Centricity Universal Viewer-00043766, M44997-Centricity Universal Viewer-00397920, M42176-Centricity Universal Viewer-00043763, A91901-Centricity Universal Viewer-01509889, 339734-Centricity PACS-IW-01444878, 283134-Centricity PACS-01265050, 282927-Centricity Universal Viewer-01806769, 2321223-Centricity Universal Viewer-01476150, 571064-Centricity Universal Viewer-01293639, 268882-Centricity Universal Viewer-01422969, 190171-Centricity Universal Viewer-00044015, 300013-Centricity Universal Viewer-00156933, AH4325-Centricity Universal Viewer-00160202, AH0116-Centricity Universal Viewer-00160203, AH0106-Centricity Universal Viewer-00160204, AP6264-Centricity Universal Viewer-00000180, 102359-1-Centricity Universal Viewer-00558696, 4306-1-Centricity Universal Viewer-00558573, 124774-Centricity Universal Viewer-00110991, RU1107-Centricity Universal Viewer-01273207, and HC0529-Centricity Universal Viewer-01364775

Distribution pattern

Worldwide Distribution - US Nationwide in the states of AK, AZ, CA, CO, DE, FL, GA, HI, IA, ID, IL, LA, MA, MN, MO, MS, MT, NC, NE, NH, NJ, NM, NV, NY, OK, OR. PA. SD, TN, TX, UT, WA, WI, WV, and WY and the countries of Austria, Belgium, Brazil, Canada, Colombia, France, Germany, Ghana, India, Ireland, Israel, Italy, Kuwait, Malaysia, Mexico, Netherlands, Panama, Poland, Russia, Saudi Arabia, South Africa, Spain, Switzerland, Turkey, United Arab Emirates, United Kingdomand Uruguay

Field note

Send feedback

We'll only use this to respond to your feedback.