Skip to content
Recall Observatory FDA recall evidence

Exact FDA event identifier

Event 79336

1 recalled-product record grouped only because every row carries this exact official event ID.

Open FDA event record

Event summary

Timeline bucket
December 15, 2017
Product types
Device
Classifications
Class II
Statuses
Terminated
Recalling firm wording
Trumpf Medical Systems, Inc.

Dossier provenance

Source snapshots represented here

  • openFDA Device Enforcement Report-date coverage June 20, 2012–July 08, 2026 Known gap: openFDA describes 2004-present coverage, but the current export has no report-date records in this interval.
  • openFDA Device Recall event_date_initiated coverage June 01, 1997–June 30, 2026

Enforcement coverage uses report dates; product initiation dates can precede those bounds. Device-enrichment coverage uses FDA event_date_initiated.

How coverage and grouping work

Complete imported group

Every recalled product in this event

1 official enforcement rows

Product dates, classifications, firms, and source wording remain attached to their individual rows. Opening this dossier never hides products that did not match a prior timeline filter.

device · product 1 of 1

Camera System, Television, Surgical without audio Intended for transmission and documentation of surgical interventions for the purpose of archiving and teaching.

Z-0824-2018
Recall number
Z-0824-2018
Initiated
December 15, 2017
Classification
Class II
Status
Terminated
Quantity
212 devices

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
The radiated emission from the TruVidia" Wireless Receiver is above the normative accepted limits, in which the radiated emission may interact with other devices in the operating room and cause them not to function properly.

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Sourced

Other

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

The radiated emission from the TruVidia" Wireless Receiver is above the normative accepted limits, in which the radiated emission may interact with other devices in the operating room and cause them not to function properly.

Code information

Catalog Number 1940747, All serial numbers: 102584474, 102584475, 102584477, 102584478, 102805330, 102813656, 102813657, 102813658, 102813659, 102813660, 102813661, 102813662, 102860647, 102860652, 102863706, 102863707, 102867783, 102867784, 102925734, 102925735, 102925736, 102925737, 102925738, 102925739, 102925740, 102925741, 102925742, 102925743, 102928703, 102932185, 102932186, 102932187, 102932188, 102932189, 102932203, 102932204, 102932205, 102932206, 102932207, 102932208, 102932209, 102932210, 102932211, 102932212, 102932217, 102932218, 102932219, 102932220, 102932221, 102932223, 102934525, 102934526, 102934527, 102934528, 102934529, 102936117, 102939484, 102940382, 102940383, 102940384, 102940386, 102940387, 102940388, 102940389, 102940390, 102940391, 102940408, 102940409, 102940410, 102961390, 102961391, 102961392, 102961393, 102961394, 102961395, 102961396, 102961397, 102961398, 102961399, 102961400, 102961401, 102975336, 102986494, 102986495, 102988480, 102989096, 102989607, 102989608, 102989615, 102990285, 102990286, 102994502, 102994503, 102994504, 102994505, 102994506, 102994525, 102994526, 102994529, 102994530, 102996306, 103005879, 103005880, 103005889, 103005890, 103005891, 103005892, 103005893, 103006636, 103010023, 103010759, 103010760, 103010761, 103010762, 103010763, 103012786, 103012787, 103013644, 103017040, 103017041, 103019900, 103020713, 103022348, 103022349, 103022350, 103022351, 103022352, 103022794, 103024049, 103024050, 103030284, 103031623, 103034106, 103034109, 103038842, 103038843, 103038844, 103038845, 103038846, 103038847, 103038848, 103038849, 103038850, 103038853, 103038854, 103038855, 103038856, 103038857, 103038858, 103038859, 103038860, 103038861, 103038862, 103038863, 103038864, 103038865, 103038866, 103039468, 103042319, 103042320, 103042321, 103042322, 103042925, 103042926, 103046119, 103046120, 103046121, 103046122, 103046123, 103046124, 103046125, 103046126, 103046127, 103046128, 103046129, 103046130, 103048248, 103065061, 103070795, 103070796, 103070797, 103070798, 103079001, 103079053, 103087365, 103087366, 103087367, 103087368, 103087369, 103087370, 103087371, 103090135, 103090136, 103090137, 103091507, 103091508, 103091509, 103091510, 103091511, 103095664, 103101676, 103101826, 103101864, 103101886, 103103841, 103103842, 103103843, 103103844, 103103845, 103104556, 103113059, 103117961, 103118986, 103118987, 103118988, 103118989, 103118990, 103118991, 103118992, 103118993, 103118994, 103118995, 103123759, 103133224, 103134656, 103138822, 103143214, 103149993, 103164298, 103168053, 103168341, 103168431, 103176559, 103176571, 103176574, 103191728, 103191729, 103194419, 103214005, 103214022.

Distribution pattern

Worldwide Distribution - US Distribution to the states of : GA, IL, IN, PA, SC, TX, UT, VA. and to the countries of : Austria, Australia, Bangladesh, Belgium, France, Switzerland, Czech Republic, Germany, Denmark, Spain, United Arab Emirates, United Kingdom, Ireland, Italy, Kuwait, Lithuania, Poland, Portugal, Romania, Sweden, Slovenia, Thailand, Turkey.

Field note

Send feedback

We'll only use this to respond to your feedback.