Event 79182

1 recalled-product record grouped only because every row carries this exact official event ID.

Open FDA event record

Event summary

Timeline bucket
November 08, 2017
Product types
Device
Classifications
Class II
Statuses
Terminated
Recalling firm wording
Med Tec Inc

Dossier provenance

Source snapshots represented here

  • openFDA Device Enforcement Report-date coverage June 20, 2012–July 08, 2026 Known gap: openFDA describes 2004-present coverage, but the current export has no report-date records in this interval.
  • openFDA Device Recall event_date_initiated coverage June 01, 1997–June 30, 2026

Enforcement coverage uses report dates; product initiation dates can precede those bounds. Device-enrichment coverage uses FDA event_date_initiated.

How coverage and grouping work

Complete imported group

Every recalled product in this event

1 official enforcement rows

Product dates, classifications, firms, and source wording remain attached to their individual rows. Opening this dossier never hides products that did not match a prior timeline filter.

device · product 1 of 1

Protura Software which utilizes Elekta's iCOM interface: MT6XSM1.4.0, MT6XSM1.4.0-1, MT6XSM1.4.0-2, MT6SXM1.4.0-3, MT6XSM1.5.0-2, MT6XSM1.6.0-1, MT6XSM1.7.2, MT6XSM1.7.2-1, MT6XSM1.7.2-3, MT6XSM1.7.3, MT6XSM1.7.3-1, MT6XSM1.7.3-3. The Protura Software is intended to interface between record and verify systems, linear accelerator (Linac) software systems, and/or image guidance systems and the Protura Couch. The system is intended to support a treatment table and add pitch and roll to normal X, Y, Z and yaw motions to provide 6-degree axis movement. The product includes an interface between the treatment planning system and the treatment table to direct the table motion based on coordinates supplied. These products connect or support a treatment table by use of various mounting plates.

Z-0671-2018
Recall number
Z-0671-2018
Initiated
November 08, 2017
Classification
Class II
Status
Terminated
Recalling firm
Med Tec Inc
Quantity
5 installations

App-derived interpretation

Device software or design reason.device_software_design · v1.0.0
software with the Elekta pedestal coordinates, moving the pedestal could cause

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Sourced

Other

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

When an error message remains displayed and is not cleared in the Protura software with the Elekta pedestal coordinates, moving the pedestal could cause the Protura software to not update with the pedestal location and buffers the pedestal movement history.

Code information

MT6XSM1.4.0, MT6XSM1.4.0-1, MT6XSM1.4.0-2, MT6SXM1.4.0-3, MT6XSM1.5.0-2, MT6XSM1.6.0-1, MT6XSM1.7.1 (M881920, M880610), MT6XSM1.7.2 (Lot M924170), MT6XSM1.7.2-1, MT6XSM1.7.2-3, MT6XSM1.7.3 (Lot M093990), MT6XSM1.7.3-1, MT6XSM1.7.3-3

Distribution pattern

US Distribution to Iowa and South Dakota. Shipped internationally to the Netherlands.