device · product 1 of 1
i-STAT DE handheld data processing module for clinical use, Software Version 2.8, List Number:08K46-01115200 (UDI: 00054749001255)
- Recall number
- Z-0926-2018
- Initiated
- September 12, 2017
- Classification
- Class II
- Status
- Terminated
- Recalling firm
- Abbott Point Of Care Inc.
- Quantity
- 894
App-derived interpretation
software version 2.8: (1) Location, operator, stored patient lists will
Official device-enrichment evidence · Sourced
Device Design
Inspect official wording and provenance
Reason for recall
Issues resulting from upgrade to software version 2.8: (1) Location, operator, stored patient lists will not update, and (2) Customized Reference Ranges, Action Ranges, and Custom Reportable Ranges are reset to factory default values. No erroneous results are generated as a result of this issue.
Code information
All units that have been upgraded to software version 2.8
Distribution pattern