Skip to content
Recall Observatory FDA recall evidence

Exact FDA event identifier

Event 78802

1 recalled-product record grouped only because every row carries this exact official event ID.

Open FDA event record

Event summary

Timeline bucket
June 15, 2017
Product types
Device
Classifications
Class II
Statuses
Terminated
Recalling firm wording
TEI Biosciences

Dossier provenance

Source snapshots represented here

  • openFDA Device Enforcement Report-date coverage June 20, 2012–July 08, 2026 Known gap: openFDA describes 2004-present coverage, but the current export has no report-date records in this interval.
  • openFDA Device Recall event_date_initiated coverage June 01, 1997–June 30, 2026

Enforcement coverage uses report dates; product initiation dates can precede those bounds. Device-enrichment coverage uses FDA event_date_initiated.

How coverage and grouping work

Complete imported group

Every recalled product in this event

1 official enforcement rows

Product dates, classifications, firms, and source wording remain attached to their individual rows. Opening this dossier never hides products that did not match a prior timeline filter.

device · product 1 of 1

PriMatrix¿ is an acellular dermal tissue matrix derived from fetal bovine dermis. The device is supplied sterile in a variety of sizes to be trimmed by the surgeon to meet the individual patient's needs.

Z-0383-2018
Recall number
Z-0383-2018
Initiated
June 15, 2017
Classification
Class II
Status
Terminated
Recalling firm
TEI Biosciences
Quantity
4081 units

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
During Final Assembly Operation, it was noticed that some of the PriMatrix Instructions for Use (IFUs) from Lot RM1608011 Part 520-002v10 contained extra pages and/or missing pages: " 22 IFUs contained duplicate pages 15 - 26 & 51 - 62 and were also missing pages 27 - 50 " 3 IFUs were missing pages 27 - 50 " 1 IFU was missing pages 1 - 2 & 75 - 76 " 1 IFU was missing pages 15 - 62 While pages 27  50, 75  76, and 15  62 contain instructions written in a language other than English, pages 1  2 contain information regarding the indications for use, contraindications, device description, warning and precautions, and potential complications of the device.

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Sourced

Error in labeling

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

During Final Assembly Operation, it was noticed that some of the PriMatrix Instructions for Use (IFUs) from Lot RM1608011 Part 520-002v10 contained extra pages and/or missing pages: " 22 IFUs contained duplicate pages 15 - 26 & 51 - 62 and were also missing pages 27 - 50 " 3 IFUs were missing pages 27 - 50 " 1 IFU was missing pages 1 - 2 & 75 - 76 " 1 IFU was missing pages 15 - 62 While pages 27  50, 75  76, and 15  62 contain instructions written in a language other than English, pages 1  2 contain information regarding the indications for use, contraindications, device description, warning and precautions, and potential complications of the device.

Code information

Product Model Number Product Lot Number Product Description Product Model Number Product Lot Number Product Description 607-004-880 1508019 8 x 8 cm 607-005-014 1702023 14mm Disc 607-005-880 1509051 8 x 8 cm 607-005-018 1702024 18mm Disc 607-005-440 1512021 4 x 4 cm 607-005-220 1702025 2 x 2 cm 607-005-812 1512022 8 x 12 cm 607-005-014 1702026 14mm Disc 607-005-440 1603044 4 x 4 cm 607-005-018 1702027 18mm Disc 607-005-660 1606038 6 x 6 cm 607-005-220 1702028 2 x 2 cm 607-001-112 1611044 10 x 12 cm 607-005-014 1702029 14mm Disc 607-001-812 1611045 8 x12 cm 607-005-018 1702030 18mm Disc 607-001-225 1612013 20 x 25 cm 607-005-220 1702031 2 x 2 cm 607-001-009 1612018 0.2 x 26.5 cm 607-005-014 1702032 14mm Disc 607-004-440 1701034 4 x 4 cm 607-005-014 1702033 14mm Disc 607-001-440 1701036 4 x 4 cm 607-005-014 1702034 14mm Disc 607-004-660 1701037 6 x 6 cm 607-005-014 1703007 14mm Disc 607-001-660 1701044 6 x 6 cm 607-005-018 1703008 18mm Disc The affected lots are manufactured from February 2017 to May 2017. The affected lots will expire in 2020 to 2021.