Monarch Airway Clearance System, Model 1000, Product Numbers: PMACS1NA (has blue tooth capability only) and PMACS1WI (has WIFI capability). The firm name on the label is Hill-Rom Services Pte. Ltd., Singapore. The Monarch Airway Clearance System, Model 1000, is intended to provide Airway Clearance Therapy and promote bronchial drainage where external manipulation of the thorax is the physicians choice of treatment. It is indicated for patients having difficulty with secretion clearance or the presence of atelectasis caused by mucus plugging.
- Recall number
- Z-0192-2018
- Initiated
- September 15, 2017
- Classification
- Class II
- Status
- Terminated
- Recalling firm
- HILL-ROM MANUFACTURING, INC.
- Quantity
- 60 devices
App-derived interpretation
Monarch Electromagnetic Interference (EMI) may disrupt the operation of implanted pacemakers or implantable cardioverter defibrillators as well as other devices that might be subject to an EMI if they come into direct contact with the Monarch device.
Official device-enrichment evidence · Sourced
Other
Inspect official wording and provenance
Reason for recall
Monarch Electromagnetic Interference (EMI) may disrupt the operation of implanted pacemakers or implantable cardioverter defibrillators as well as other devices that might be subject to an EMI if they come into direct contact with the Monarch device.
Code information
Serial numbers S082VM0092, S122VM0171, S128VM0205, S151VM0242, S082VM0093, S122VM0172, S132VM0210, S151VM0243, S082VM0095, S122VM0175, S132VM0212, S152VM0251, S095VM0119, S122VM0176, S132VM0213, S157VM0257, S095VM0124, S124VM0177, S135VM0216, S157VM0261, S095VM0125, S124VM0178, S136VM0219, S158VM0264, S114VM0152, S124VM0180, S136VM0222, S158VM0266, S119VM0158, S124VM0181, S136VM0223, S158VM0267, S119VM0160, S126VM0187, S138VM0224, S158VM0268, S119VM0161, S126VM0193, S138VM0226, S158VM0271, S119VM0162, S126VM0198, S146VM0227, S158VM0273, S122VM0165, S126VM0199, S146VM0229, S159VM0276, S122VM0166, S128VM0202, S146VM0230, S160VM0279, S122VM0168, S128VM0203, S151VM0238, S167VM0289, S122VM0170, S128VM0204, S151VM0240, and S167VM0291.
Distribution pattern
USA (nationwide) Distribution to the states of: AZ, CA, CO, CT, FL, GA, IA, IL, IN, KY, MA, MI, MN, MO, NE, OH, OK, OR, TN, TX, VA, WA, WI, and WV. Internationally to Canada. There was no military/government distribution.