Event summary

Timeline bucket
September 07, 2017
Product types
Device
Classifications
Class II
Statuses
Terminated
Recalling firm wording
Radiometer America Inc

Dossier provenance

Source snapshots represented here

  • openFDA Device Enforcement Report-date coverage June 20, 2012–July 08, 2026 Known gap: openFDA describes 2004-present coverage, but the current export has no report-date records in this interval.
  • openFDA Device Recall event_date_initiated coverage June 01, 1997–June 30, 2026

Enforcement coverage uses report dates; product initiation dates can precede those bounds. Device-enrichment coverage uses FDA event_date_initiated.

How coverage and grouping work

Complete imported group

Every recalled product in this event

1 official enforcement rows

Product dates, classifications, firms, and source wording remain attached to their individual rows. Opening this dossier never hides products that did not match a prior timeline filter.

device · product 1 of 1

AutoCheck5+, Level 2, specifically designed for Radiometer Americas blood gas analyzers The QUALICHECK+ quality control (QC) ampoules are specifically designed for Radiometer America s blood gas analyzers

Z-3237-2017
Recall number
Z-3237-2017
Initiated
September 07, 2017
Classification
Class II
Status
Terminated
Recalling firm
Radiometer America Inc
Quantity
1669 Boxes

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
Radiometer has recently become aware that a few of the AUTOCHECK 5+ level 2 ampoules, will give incorrect results for pCO2 and pO2. A defective QC ampoule will for: " pCO2 give a result close to or just outside of the low end of the control range. (4.92  5.85 kPa or 36.9  46.9 mmHg) " pO2 give a considerable higher result than the expected assigned value of 13.9 -14.3 kPa (104-107 mmHg)

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Unknown

An exact joined enrichment record exists, but none supplies supported root-cause wording. This is not an FDA finding of an unknown cause.

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

Radiometer has recently become aware that a few of the AUTOCHECK 5+ level 2 ampoules, will give incorrect results for pCO2 and pO2. A defective QC ampoule will for: " pCO2 give a result close to or just outside of the low end of the control range. (4.92  5.85 kPa or 36.9  46.9 mmHg) " pO2 give a considerable higher result than the expected assigned value of 13.9 -14.3 kPa (104-107 mmHg)

Code information

S7745 Lot R0698

Distribution pattern

Worldwide Distribution - US (nationwide) Internationally to Canada, Netherlands, China, Czech Republic, Denmark, Germany, Spain France, Hungary, India, Denmark, Tokyo, South Korea, United Kingdom, Australia, Poland, South Africa, Singapore, Switzerland, Turkey, Norway, Austria, Italy, Sweden, and Finland.

Field note

Send feedback

We'll only use this to respond to your feedback.