device · product 1 of 1
AutoCheck5+, Level 2, specifically designed for Radiometer Americas blood gas analyzers The QUALICHECK+ quality control (QC) ampoules are specifically designed for Radiometer America s blood gas analyzers
- Recall number
- Z-3237-2017
- Initiated
- September 07, 2017
- Classification
- Class II
- Status
- Terminated
- Recalling firm
- Radiometer America Inc
- Quantity
- 1669 Boxes
App-derived interpretation
Radiometer has recently become aware that a few of the AUTOCHECK 5+ level 2 ampoules, will give incorrect results for pCO2 and pO2. A defective QC ampoule will for: " pCO2 give a result close to or just outside of the low end of the control range. (4.92 5.85 kPa or 36.9 46.9 mmHg) " pO2 give a considerable higher result than the expected assigned value of 13.9 -14.3 kPa (104-107 mmHg)
Official device-enrichment evidence · Unknown
An exact joined enrichment record exists, but none supplies supported root-cause wording. This is not an FDA finding of an unknown cause.
Inspect official wording and provenance
Reason for recall
Radiometer has recently become aware that a few of the AUTOCHECK 5+ level 2 ampoules, will give incorrect results for pCO2 and pO2. A defective QC ampoule will for: " pCO2 give a result close to or just outside of the low end of the control range. (4.92 5.85 kPa or 36.9 46.9 mmHg) " pO2 give a considerable higher result than the expected assigned value of 13.9 -14.3 kPa (104-107 mmHg)
Code information
S7745 Lot R0698
Distribution pattern
Worldwide Distribution - US (nationwide) Internationally to Canada, Netherlands, China, Czech Republic, Denmark, Germany, Spain France, Hungary, India, Denmark, Tokyo, South Korea, United Kingdom, Australia, Poland, South Africa, Singapore, Switzerland, Turkey, Norway, Austria, Italy, Sweden, and Finland.