device · product 1 of 1
Arctic Sun 5000 Temperature Management System, 100-120 V North America and 100-120 V US
- Recall number
- Z-3042-2017
- Initiated
- March 31, 2017
- Classification
- Class II
- Status
- Ongoing
- Recalling firm
- C.R. Bard, Inc.
- Quantity
- 3,450 units
App-derived interpretation
Potential failure of the Arctic Sun 5000 Temperature Management System may occur when the continued use is in excess of 2000 system hours without preventive maintenance.
Official device-enrichment evidence · Unknown
An exact joined enrichment record exists, but none supplies supported root-cause wording. This is not an FDA finding of an unknown cause.
Inspect official wording and provenance
Reason for recall
Potential failure of the Arctic Sun 5000 Temperature Management System may occur when the continued use is in excess of 2000 system hours without preventive maintenance.
Code information
Product Codes 5000-00-00 & 5000-00-00E - All serial numbers
Distribution pattern
Worldwide Distribution - USA (nationwide) to the states of :AL, AK, AZ, AR, CA, CO, CT, DE, DC, FL, GA, HI, ID, IL, IN, IA, KS, KY, LA, ME, MD, MA, MI, MN, MS MO, MT, NE, NV, NH, NJ, NM, NY, NC, ND, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VT, VA, WA, WV, WY., and to the countries of : Australia, Korea, Japan, Canada, Hong Kong, China, Singapore, & EMEA