Skip to content
Recall Observatory FDA recall evidence

Exact FDA event identifier

Event 77195

1 recalled-product record grouped only because every row carries this exact official event ID.

Open FDA event record

Event summary

Timeline bucket
February 01, 2017
Product types
Device
Classifications
Class II
Statuses
Terminated
Recalling firm wording
Toshiba American Medical Systems Inc

Dossier provenance

Source snapshots represented here

  • openFDA Device Enforcement Report-date coverage June 20, 2012–July 08, 2026 Known gap: openFDA describes 2004-present coverage, but the current export has no report-date records in this interval.
  • openFDA Device Recall event_date_initiated coverage June 01, 1997–June 30, 2026

Enforcement coverage uses report dates; product initiation dates can precede those bounds. Device-enrichment coverage uses FDA event_date_initiated.

How coverage and grouping work

Complete imported group

Every recalled product in this event

1 official enforcement rows

Product dates, classifications, firms, and source wording remain attached to their individual rows. Opening this dossier never hides products that did not match a prior timeline filter.

device · product 1 of 1

Angio WorkStation (XIDF-AWS801) used in conjunction with your Infinix System (INFX-8000V;INFX-8000C;INFX-8000F

Z-2109-2017
Recall number
Z-2109-2017
Initiated
February 01, 2017
Classification
Class II
Status
Terminated
Quantity
US - 70 systems

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
It was found that during a procedure the Peak Skin Dose (PSD) value displayed by the Dose Tracking System (DTS) was larger than the DTS expected value. This issue occurs when the power for the Angio Workstation is turned on after the power for the Infinix system is turned on.

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Sourced

Radiation Control for Health and Safety Act

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

It was found that during a procedure the Peak Skin Dose (PSD) value displayed by the Dose Tracking System (DTS) was larger than the DTS expected value. This issue occurs when the power for the Angio Workstation is turned on after the power for the Infinix system is turned on.

Code information

S/N : 8181622393

Distribution pattern

US Distribution to the states of :LA, IL, NY, FL, MA, TN, MO, GA, CA, DE, MO, AL, ID, MN, TX, CT, PA, MS, OH, NC, NV, W VA, WA, AR, NJ, MD, OK, IN and NH.

Field note

Send feedback

We'll only use this to respond to your feedback.