Skip to content
Recall Observatory FDA recall evidence

Exact FDA event identifier

Event 75898

1 recalled-product record grouped only because every row carries this exact official event ID.

Open FDA event record

Event summary

Timeline bucket
May 13, 2015
Product types
Device
Classifications
Class II
Statuses
Terminated
Recalling firm wording
The Binding Site Group, Ltd.

Dossier provenance

Source snapshots represented here

  • openFDA Device Enforcement Report-date coverage June 20, 2012–July 08, 2026 Known gap: openFDA describes 2004-present coverage, but the current export has no report-date records in this interval.
  • openFDA Device Recall event_date_initiated coverage June 01, 1997–June 30, 2026

Enforcement coverage uses report dates; product initiation dates can precede those bounds. Device-enrichment coverage uses FDA event_date_initiated.

How coverage and grouping work

Complete imported group

Every recalled product in this event

1 official enforcement rows

Product dates, classifications, firms, and source wording remain attached to their individual rows. Opening this dossier never hides products that did not match a prior timeline filter.

device · product 1 of 1

Optilite Analyser, Product Code: IE700 Product Usage: The Thermo Scientific Indiko Clinical Chemistry Analyzer is a fully automated random access analyzer used to measure a variety of analytes that may be adaptable to the analyzer depending on the reagent used. The Indiko Glucose (H-K) test system, is intended for in vitro diagnostic use in the quantitative determination of the glucose concentration in human plasma on the Indiko analyzer. Glucose measurements are used in the diagnosis and treatment of carbohydrate metabolism disorders including diabetes mellitus, neonatal hypoglycemia, and idiopathic hypoglycemnia, and of pancreatic islet cell carcinoma.

Z-1018-2017
Recall number
Z-1018-2017
Initiated
May 13, 2015
Classification
Class II
Status
Terminated
Quantity
129 units

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
Potential risk of the Optilite lid/cover falling suddenly and causing injury when not placed in the fully open position.

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Sourced

Component design/selection

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

Potential risk of the Optilite lid/cover falling suddenly and causing injury when not placed in the fully open position.

Code information

864000290243 864000290832 864000290244 864000290833 864000290330 864000290834 864000290447 864000290836 864000290448 864000290837 864000290449 864000290838 864000290452 864000290844 864000290826 864000290845 864000290827 864000290846 864000290828 864000290847 864000290831 864000290850

Distribution pattern

Worldwide Distribution - US Nationwide in the states of: CA, VA, GA, PA, NJ, KY, NY, AR, NM, NC, AL, GA, MI, and the countries of Canada, China, Australia, Israel, Ireland, UK, Germany, Belgium, Italy, France, Czech republic, Spain, Portugal.

Field note

Send feedback

We'll only use this to respond to your feedback.