device · product 1 of 1
DELFIA hFSH kit. In Vitro Diagnostic for the determination of hFSH in serum
A DELFIA hFSH kit lot failed real time stability testing. The real time stability data demonstrated a decrease in hFSH concentrations between -4.8% to -11.3% for control sera. The -11.3% bias was observed at an hFSH target concentration of 101 IU/L. The change observed in the serum samples within the range 1.76 67.0 IU/L hFSH averaged -2.9%. The reference range of the assay spans 0.60 to 98.0 IU/L.
These labels are deterministic app interpretations, not FDA categories.
A DELFIA hFSH kit lot failed real time stability testing. The real time stability data demonstrated a decrease in hFSH concentrations between -4.8% to -11.3% for control sera. The -11.3% bias was observed at an hFSH target concentration of 101 IU/L. The change observed in the serum samples within the range 1.76 67.0 IU/L hFSH averaged -2.9%. The reference range of the assay spans 0.60 to 98.0 IU/L.
Code information
Lot Number: 646091
Distribution pattern
CA, CO, NY, WA Foreign: Austria, Bolivia, Bosnia , Sweden, Norway, Netherlands, Japan, Great Britain,