Event 75240

1 recalled-product record grouped only because every row carries this exact official event ID.

Open FDA event record

Event summary

Timeline bucket
September 09, 2016
Product types
Device
Classifications
Class II
Statuses
Terminated
Recalling firm wording
Contour Fabricators Inc

Dossier provenance

Source snapshots represented here

  • openFDA Device Enforcement Report-date coverage June 20, 2012–July 08, 2026 Known gap: openFDA describes 2004-present coverage, but the current export has no report-date records in this interval.
  • openFDA Device Recall event_date_initiated coverage June 01, 1997–June 30, 2026

Enforcement coverage uses report dates; product initiation dates can precede those bounds. Device-enrichment coverage uses FDA event_date_initiated.

How coverage and grouping work

Complete imported group

Every recalled product in this event

1 official enforcement rows

Product dates, classifications, firms, and source wording remain attached to their individual rows. Opening this dossier never hides products that did not match a prior timeline filter.

device · product 1 of 1

Zero-Gravity Radiation Protection System Radiology Personnel protective shield.

Z-0080-2017
Recall number
Z-0080-2017
Initiated
September 09, 2016
Classification
Class II
Status
Terminated
Recalling firm
Contour Fabricators Inc
Quantity
447

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
Two customer complaints suggest that excessive rotation of the cable in one direction (clockwise or counter-clockwise) without allowing the cable to unwind to its normal condition may cause the cable to break suddenly and the body shield to fall. While no injuries or deaths occurred in either failure, the Zero-GravityTM body shield weighs 55 pounds, and in the event of a cable failure, the body shield will fall immediately. A fall of this nature places the user, associated staff and the patient at risk for a potential injury.

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Sourced

Labeling design

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

Two customer complaints suggest that excessive rotation of the cable in one direction (clockwise or counter-clockwise) without allowing the cable to unwind to its normal condition may cause the cable to break suddenly and the body shield to fall. While no injuries or deaths occurred in either failure, the Zero-GravityTM body shield weighs 55 pounds, and in the event of a cable failure, the body shield will fall immediately. A fall of this nature places the user, associated staff and the patient at risk for a potential injury.

Code information

All serial numbers for all models listed below manufactured/distributed between June 28, 2010  August 25, 2016. Part No. : Description ZGM-6-.5H : Zero-Gravity floor unit ZGM-6-5ROUNDH : Zero-Gravity floor unit ZGM-6Demo : Zero-Gravity demo floor unit ZGCM-48 : Zero-Gravity ceiling mounted monorail with 48 boom arm ZGCM-48-SPECIAL : Zero-Gravity ceiling mounted monorail with 48 boom arm with monorail extension ZG48 : Zero-Gravity 48 boom arm assembly and body shield for monorail ZGCM-66 : Zero-Gravity ceiling mounted monorail with 66 boom arm ZGCM-HSA : Zero-Gravity ceiling mounted monorail with hinged swing arm ZGCH : Zero-Gravity ceiling mounted monorail Contour track with 48 boom arm ZGCH48 : Zero-Gravity 48 boom arm assembly and body shield for Contour track ZGHSA : Zero-Gravity single point, ceiling mount with hinged swing arm ZGCSP : Zero-Gravity single point, ceiling mount with hinged swing arm 29858 : Zero-Gravity floor unit

Distribution pattern

Nationwide. Canada. AUSTRALIA AUSTRIA BELGIUM BRUNEI CHINA FRANCE GERMANY INDA INDONESIA ISRAEL ITALY KSA MALAYSIA NETHERLANDS POLAND QATAR QUEBEC SAUDI ARABIA SINGAPORE SOUTH AFRICA TAIWAN (R.O.C.)