Skip to content
Recall Observatory FDA recall evidence

Exact FDA event identifier

Event 75213

1 recalled-product record grouped only because every row carries this exact official event ID.

Open FDA event record

Event summary

Timeline bucket
September 08, 2016
Product types
Device
Classifications
Class II
Statuses
Terminated
Recalling firm wording
Argon Medical Devices, Inc

Dossier provenance

Source snapshots represented here

  • openFDA Device Enforcement Report-date coverage June 20, 2012–July 08, 2026 Known gap: openFDA describes 2004-present coverage, but the current export has no report-date records in this interval.
  • openFDA Device Recall event_date_initiated coverage June 01, 1997–June 30, 2026

Enforcement coverage uses report dates; product initiation dates can precede those bounds. Device-enrichment coverage uses FDA event_date_initiated.

How coverage and grouping work

Complete imported group

Every recalled product in this event

1 official enforcement rows

Product dates, classifications, firms, and source wording remain attached to their individual rows. Opening this dossier never hides products that did not match a prior timeline filter.

device · product 1 of 1

MicroStic Plus w/SS Wire 5F and MicroStic Plus w/NiTi Wire & Echo Needle 4F. A sheath used to facilitate placing a catheter through the skin into a vein or artery.

Z-0051-2017
Recall number
Z-0051-2017
Initiated
September 08, 2016
Classification
Class II
Status
Terminated
Quantity
200 units

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
The supplier initiated a recall because affected lots of the micro-introducer have an outer diameter of the dilator hub that is slightly smaller than is capable of securing the dilator to the sheath

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Sourced

Nonconforming Material/Component

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

The supplier initiated a recall because affected lots of the micro-introducer have an outer diameter of the dilator hub that is slightly smaller than is capable of securing the dilator to the sheath

Code information

11149662, 11151447, 11151617, and 11149822.

Distribution pattern

Worldwide Distribution -- US, to the states of TN and MS; and the country of Canada.

Field note

Send feedback

We'll only use this to respond to your feedback.