device · product 1 of 1
CDI¿ H/S Cuvette The intended use for the CDI¿ H IS Cuvette is as follows: The CDI¿ H/S Cuvettes are intended for use with a CDI monitor during cardiopulmonary bypass procedures when continuous monitoring of blood hematocrit and oxygen saturation is desired.
Terumo Cardiovascular Systems (Terumo CVS) has received complaints of the CDI System 500 monitor displaying the H/S DISCONNECT AT CUVETTE error message when the CDI H/S Cuvette does not make a proper connection to the CDI H/S Probe. In this situation, blood parameter values for HCT, Hgb, and SO2 do not display.
These labels are deterministic app interpretations, not FDA categories.
Terumo Cardiovascular Systems (Terumo CVS) has received complaints of the CDI System 500 monitor displaying the H/S DISCONNECT AT CUVETTE error message when the CDI H/S Cuvette does not make a proper connection to the CDI H/S Probe. In this situation, blood parameter values for HCT, Hgb, and SO2 do not display.