device · product 1 of 1
Hitachi Scenaria Computed Tomography System Hitachi Medical Systems America, Twinsburg, OH 44087
The firm discovered that the centrifugal force applied to the power supply may cause it to fail over time. The exam performed at the time of the failure will likely be lost, resulting in the need to rescan the patient.
These labels are deterministic app interpretations, not FDA categories.
The firm discovered that the centrifugal force applied to the power supply may cause it to fail over time. The exam performed at the time of the failure will likely be lost, resulting in the need to rescan the patient.
Code information
Serial #'s: S5002-S5014; S5017-S5038
Distribution pattern
US Distribution to the states of : CA, FL, GA, IA, IL, IN, MD, NC, NY, OH, PA, SD, TN, TX & WY., and Internationally to Brazil and Mexico.