Skip to content
Recall Observatory FDA recall evidence

Exact FDA event identifier

Event 74341

1 recalled-product record grouped only because every row carries this exact official event ID.

Open FDA event record

Event summary

Timeline bucket
May 09, 2016
Product types
Device
Classifications
Class II
Statuses
Terminated
Recalling firm wording
Datascope Corporation

Dossier provenance

Source snapshots represented here

  • openFDA Device Enforcement Report-date coverage June 20, 2012–July 08, 2026 Known gap: openFDA describes 2004-present coverage, but the current export has no report-date records in this interval.
  • openFDA Device Recall event_date_initiated coverage June 01, 1997–June 30, 2026

Enforcement coverage uses report dates; product initiation dates can precede those bounds. Device-enrichment coverage uses FDA event_date_initiated.

How coverage and grouping work

Complete imported group

Every recalled product in this event

1 official enforcement rows

Product dates, classifications, firms, and source wording remain attached to their individual rows. Opening this dossier never hides products that did not match a prior timeline filter.

device · product 1 of 1

Small Patient ECC Pack, REF BEQ-TOP 24100, Product Usage: The HLM Tubing Sets with Bioline Coating are for single use only. They may be sold sterile, non-sterile and bulk packed. Custom tubing sets that are sold sterile are not to be re-sterilized by the user. In open heart surgery the HLM Tubing Sets with Bioline Coating are used in combination with the heart-lung machine for the oxygenation of blood and removal of carbon dioxide. The main purpose of the HLM Tubing Sets with Bioline Coating is to connect the patient to the hear-lung machine and its components. The HLM Tubing Sets with Bioline Coating are therefore a component in the extracorporeal perfusion circulation system. The utilization period of the use of the tubing sets is restricted to six hours

Z-2002-2016
Recall number
Z-2002-2016
Initiated
May 09, 2016
Classification
Class II
Status
Terminated
Recalling firm
Datascope Corporation
Quantity
17 units

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
Custom Tubing Set BEQ-TOP-24100 (Small Patient ECC Pack) was identified as possessing a level of endotoxin that exceeded the regulatory limits.

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Unknown

An exact joined enrichment record exists, but none supplies supported root-cause wording. This is not an FDA finding of an unknown cause.

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

Custom Tubing Set BEQ-TOP-24100 (Small Patient ECC Pack) was identified as possessing a level of endotoxin that exceeded the regulatory limits.

Code information

Lot/batch No: 3000003260, Part number 701050175

Distribution pattern

US in the state of GA

Field note

Send feedback