Skip to content
Recall Observatory FDA recall evidence

Exact FDA event identifier

Event 74114

1 recalled-product record grouped only because every row carries this exact official event ID.

Open FDA event record

Event summary

Timeline bucket
May 10, 2016
Product types
Device
Classifications
Class II
Statuses
Terminated
Recalling firm wording
Integra LifeSciences Corp.

Dossier provenance

Source snapshots represented here

  • openFDA Device Enforcement Report-date coverage June 20, 2012–July 08, 2026 Known gap: openFDA describes 2004-present coverage, but the current export has no report-date records in this interval.
  • openFDA Device Recall event_date_initiated coverage June 01, 1997–June 30, 2026

Enforcement coverage uses report dates; product initiation dates can precede those bounds. Device-enrichment coverage uses FDA event_date_initiated.

How coverage and grouping work

Complete imported group

Every recalled product in this event

1 official enforcement rows

Product dates, classifications, firms, and source wording remain attached to their individual rows. Opening this dossier never hides products that did not match a prior timeline filter.

device · product 1 of 1

TS Series Depth Gauge for Integra Ti6 Internal Fixation System 2.0/2.5mm Screws The FS2030 Depth Gauge is used by the physician to determine the correct screw length when implanting a 2.0 or 2.5mm TS Series screw for the Ti6 Internal Fixation System.

Z-1965-2016
Recall number
Z-1965-2016
Initiated
May 10, 2016
Classification
Class II
Status
Terminated
Quantity
71 units

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
As a result of an internal investigation of a non-conformance it has been determined: 1) Depth Gauges from Lot UP15H2 exceed the overall length specification requirements. 2) Depth Gauges from prior lot numbers manufactured before Integra acquired the product line reflect an incorrect number for the first mm depth marking on the gauge. The number should be 4, it is 6. 3) The space / holder

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Sourced

Process design

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

As a result of an internal investigation of a non-conformance it has been determined: 1) Depth Gauges from Lot UP15H2 exceed the overall length specification requirements. 2) Depth Gauges from prior lot numbers manufactured before Integra acquired the product line reflect an incorrect number for the first mm depth marking on the gauge. The number should be 4, it is 6. 3) The space / holder

Code information

Lot UP15H2, Lot UP09D3, Lot UP07E2

Distribution pattern

US Distribution to states of: CO, FL, IL, MI, NJ, NY, SC, TX, and VA.

Field note

Send feedback