Skip to content
Recall Observatory FDA recall evidence

Exact FDA event identifier

Event 73172

4 recalled-product records grouped only because every row carries this exact official event ID.

Open FDA event record

Event summary

Timeline bucket
February 04, 2016
Product types
Device
Classifications
Class II
Statuses
Terminated
Recalling firm wording
Stryker Instruments Div. of Stryker Corporation

Dossier provenance

Source snapshots represented here

  • openFDA Device Enforcement Report-date coverage June 20, 2012–July 08, 2026 Known gap: openFDA describes 2004-present coverage, but the current export has no report-date records in this interval.
  • openFDA Device Recall event_date_initiated coverage June 01, 1997–June 30, 2026

Enforcement coverage uses report dates; product initiation dates can precede those bounds. Device-enrichment coverage uses FDA event_date_initiated.

How coverage and grouping work

Complete imported group

Every recalled product in this event

4 official enforcement rows

Product dates, classifications, firms, and source wording remain attached to their individual rows. Opening this dossier never hides products that did not match a prior timeline filter.

device · product 1 of 4

(Stryker) AccuPlace Straight Level Needle Guide; Assists the user with inserting a needle at a predetermined angle for computer tomography.

Z-1037-2016
Recall number
Z-1037-2016
Initiated
February 04, 2016
Classification
Class II
Status
Terminated
Quantity
5,800 units total

App-derived interpretation

Sterility assurance reason.sterility_assurance · v1.0.0
breach in sterility

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Sourced

Package design/selection

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

During an evaluation of the product packaging, it was determined that there is a potential for the product to cause damage to the packaging, creating a possible breach in sterility.

Code information

Stryker Model: 0900-300-000; With lot numbers: 61405002, 61406007, 61408002, and 61509002. Devices distributed between 5-1-2013 and 5-31-2014 are labeled with Inrads name and product number. Devices distributed between 6-20-2014 and 12-3-15 are labeled with Strykers name and product number.

Distribution pattern

Nationwide Distribution-including the states of AZ, CA, CO, CT, FL, GA, IA, KS, MA, MI, MN, MO, NY, OH, OK, PA, TX, VA, WA, and WI.

device · product 2 of 4

(Stryker) AccuPlace Round Level Needle Guide; Assists the user with inserting a needle at a predetermined angle for computer tomography.

Z-1038-2016
Recall number
Z-1038-2016
Initiated
February 04, 2016
Classification
Class II
Status
Terminated
Quantity
5,800 units total

App-derived interpretation

Sterility assurance reason.sterility_assurance · v1.0.0
breach in sterility

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Sourced

Package design/selection

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

During an evaluation of the product packaging, it was determined that there is a potential for the product to cause damage to the packaging, creating a possible breach in sterility.

Code information

Stryker Model: 0900-400-000; With lot numbers: 61406002, 61505003. Devices distributed between 5-1-2013 and 5-31-2014 are labeled with Inrads name and product number. Devices distributed between 6-20-2014 and 12-3-15 are labeled with Strykers name and product number.

Distribution pattern

Nationwide Distribution-including the states of AZ, CA, CO, CT, FL, GA, IA, KS, MA, MI, MN, MO, NY, OH, OK, PA, TX, VA, WA, and WI.

device · product 3 of 4

50000 (Inrad) AccuPlace Drace Stereotaxic Needle Guide; Assists the user with inserting a needle at a predetermined angle for computer tomography.

Z-1039-2016
Recall number
Z-1039-2016
Initiated
February 04, 2016
Classification
Class II
Status
Terminated
Quantity
5,800 units total

App-derived interpretation

Sterility assurance reason.sterility_assurance · v1.0.0
breach in sterility

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Sourced

Package design/selection

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

During an evaluation of the product packaging, it was determined that there is a potential for the product to cause damage to the packaging, creating a possible breach in sterility.

Code information

50000 Lot numbers: 61305001, 61310001, 61312001, 61404006 Devices distributed between 5-1-2013 and 5-31-2014 are labeled with Inrads name and product number. Devices distributed between 6-20-2014 and 12-3-15 are labeled with Strykers name and product number.

Distribution pattern

Nationwide Distribution-including the states of AZ, CA, CO, CT, FL, GA, IA, KS, MA, MI, MN, MO, NY, OH, OK, PA, TX, VA, WA, and WI.

device · product 4 of 4

50001 (Inrad) AccuPlace Drace Stereotaxic Needle Guide; Assists the user with inserting a needle at a predetermined angle for computer tomography.

Z-1040-2016
Recall number
Z-1040-2016
Initiated
February 04, 2016
Classification
Class II
Status
Terminated
Quantity
5,800 units total

App-derived interpretation

Sterility assurance reason.sterility_assurance · v1.0.0
breach in sterility

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Sourced

Package design/selection

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

During an evaluation of the product packaging, it was determined that there is a potential for the product to cause damage to the packaging, creating a possible breach in sterility.

Code information

50001 Lot numbers: 61303001, 61309001, 61310001, 61404006 Devices distributed between 5-1-2013 and 5-31-2014 are labeled with Inrads name and product number. Devices distributed between 6-20-2014 and 12-3-15 are labeled with Strykers name and product number.

Distribution pattern

Nationwide Distribution-including the states of AZ, CA, CO, CT, FL, GA, IA, KS, MA, MI, MN, MO, NY, OH, OK, PA, TX, VA, WA, and WI.