3M Steri-Vac Gas Sterilizer/Aerator, models 5XL and 8XL.
- Recall number
- Z-1563-2016
- Initiated
- January 22, 2016
- Classification
- Class II
- Status
- Terminated
- Recalling firm
- 3M Company - Health Care Business
- Quantity
- 74 valves
App-derived interpretation
incorrectly labeled
Official device-enrichment evidence · Sourced
Nonconforming Material/Component
Inspect official wording and provenance
Reason for recall
3M discovered a valve with incorrectly labeled "input" and "output" orientation. This defect can result in the valve being installed backwards, creating a remote possibility of valve failure. A failure may result in EO being vented outside the sealed chamber during a limited period in the EO injection phase of the cycle.
Code information
Model 5XL, Serial Numbers: 722612 - 722683, 820352 - 820360 Model 8XL, Serial Numbers: 351303 - 351399, 450144 - 450147
Distribution pattern
AL, CO, CT, FL, GA, IN, NC, NJ, NY, OK, OR, PA, RI, VA, WI, and OUS to include: Argentina, Canada, Chile, China, Columbia, El Salvador, Germany, India, Indonesia, Japan, Korea, Mexico, Peru, Philippines, South Africa, Switzerland, Thailand, Trinidad, Turkey, United Arab Emirates, Vietnam.