Skip to content
Recall Observatory FDA recall evidence

Exact FDA event identifier

Event 72879

1 recalled-product record grouped only because every row carries this exact official event ID.

Open FDA event record

Event summary

Timeline bucket
December 09, 2015
Product types
Device
Classifications
Class II
Statuses
Terminated
Recalling firm wording
Philips Ultrasound, Inc.

Dossier provenance

Source snapshots represented here

  • openFDA Device Enforcement Report-date coverage June 20, 2012–July 08, 2026 Known gap: openFDA describes 2004-present coverage, but the current export has no report-date records in this interval.
  • openFDA Device Recall event_date_initiated coverage June 01, 1997–June 30, 2026

Enforcement coverage uses report dates; product initiation dates can precede those bounds. Device-enrichment coverage uses FDA event_date_initiated.

How coverage and grouping work

Complete imported group

Every recalled product in this event

1 official enforcement rows

Product dates, classifications, firms, and source wording remain attached to their individual rows. Opening this dossier never hides products that did not match a prior timeline filter.

device · product 1 of 1

Philips Lumify Diagnostic Ultrasound, Catalogue Number: 795216 Part Number: 989605449841 with 453561845331 (software version 1.0)

Z-0596-2016
Recall number
Z-0596-2016
Initiated
December 09, 2015
Classification
Class II
Status
Terminated
Recalling firm
Philips Ultrasound, Inc.
Quantity
21 units in the US only

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
Color Flow direction is displayed incorrectly in Lumify 1.0. The system displayed Color Flow direction does not correctly represent the annotated Color Bar or Velocity Markers. (e.g.) When the Color Bar conveys that Red is to be displayed for color flow toward the Transducer, the system displays Blue for color flow toward the Transducer. This could result in misdiagnosis in some studies.

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Sourced

Software design

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

Color Flow direction is displayed incorrectly in Lumify 1.0. The system displayed Color Flow direction does not correctly represent the annotated Color Bar or Velocity Markers. (e.g.) When the Color Bar conveys that Red is to be displayed for color flow toward the Transducer, the system displays Blue for color flow toward the Transducer. This could result in misdiagnosis in some studies.

Code information

System Serial numbers: USD15L0177; USD15L0178; USD15L0179; USD15L0297; USD15L0298; USN15L0789; USN15L0862; USN15L0863; USN15L0864; USN15L0865; USN15L0866; USN15L0867; USN15L0917; USN15L0918; USN15L0988; USO15L0563; USO15L0564; USO15L0565; USO15L0566; USO15L1087; and USO15L1088.

Distribution pattern

distributed in CA, CT, ND, NV, OR, RI, TN, and WA.

Field note

Send feedback

We'll only use this to respond to your feedback.