Skip to content
Recall Observatory FDA recall evidence

Exact FDA event identifier

Event 72713

1 recalled-product record grouped only because every row carries this exact official event ID.

Open FDA event record

Event summary

Timeline bucket
November 17, 2015
Product types
Device
Classifications
Class II
Statuses
Terminated
Recalling firm wording
Olympus Corporation of the Americas

Dossier provenance

Source snapshots represented here

  • openFDA Device Enforcement Report-date coverage June 20, 2012–July 08, 2026 Known gap: openFDA describes 2004-present coverage, but the current export has no report-date records in this interval.
  • openFDA Device Recall event_date_initiated coverage June 01, 1997–June 30, 2026

Enforcement coverage uses report dates; product initiation dates can precede those bounds. Device-enrichment coverage uses FDA event_date_initiated.

How coverage and grouping work

Complete imported group

Every recalled product in this event

1 official enforcement rows

Product dates, classifications, firms, and source wording remain attached to their individual rows. Opening this dossier never hides products that did not match a prior timeline filter.

device · product 1 of 1

Olympus ENDOEYE HD II Video Telescopes. Models WA50040A ENDOEYE HD II, 10 mm, 0 degree, WA50042A ENDOEYE HD II 10 mm, 30 degree, W7100603 Repair cable 10 mm 0/30 degree. The ENDOEYE video Telescopes are used with other supporting equipment for endoscopy and endoscopic surgery within the thoracic and abdominal cavities including the female reproductive organs.

Z-0403-2016
Recall number
Z-0403-2016
Initiated
November 17, 2015
Classification
Class II
Status
Terminated
Quantity
182

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
A damaged temperature sensor at the ENDOEYE tip caused the distal end to become abnormally hot. Excessive heating of the ENDOEYE distal end could result in patient or user injury.

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Unknown

An exact joined enrichment record exists, but none supplies supported root-cause wording. This is not an FDA finding of an unknown cause.

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

A damaged temperature sensor at the ENDOEYE tip caused the distal end to become abnormally hot. Excessive heating of the ENDOEYE distal end could result in patient or user injury.

Code information

Serial numbers 622483, 622523, 622553, 622566, 622568, 622569, 622573, 622576, 622614, 622615, 622616, 622617, 622618, 622619, 622620, 622632, 622643, 622644, 622645, 622646, 622647, 622654, 622655, 622656, 622657, 622658, 622664, 622666, 622667, 622687, 622688, 622693, 622704, 622705, 622706, 622710, 622711, 622712, 622718, 622719, 622720, 622722, 622723, 622750, 622753, 622764, 622766, 622767, 622769, 622771, 622773, 622775, 622777, 622778, 622785, 622789, 622795, 622797, 622798, 622801, 622806, 623624, 623626, 623629, 623637, 623638, 623639, 623641, 623661, 623669, 623671, 623672, 623673, 623677, 623687, 623688, 623705, 623709, 623713, 623719, 623720, 623722, 623723, 623729, 623730, 623731, 623735, 623736, 623737, 623738, 623739, 623740, 623741, 623742, 623743, 623747, 623749, 623750, 623751, 623752, 623776, 623778, 623784, 623786, 623787, 623788, 623799, 623800, 623801, 623802, 624004, 624005, 624006, 624008, 624011, 624012, 624016, 624017, 624018, 624019, 624024, 624028, 624030, 624041, 624042, 624046, 624054, 624055, 624056, 624057, 624064, 624066, 624070, 624071, 624073, 624076, 624077, 624080, 624082, 624089, 624099, 624100, 624101, 624104, 624107, 624109, 624110, 624112, 624114, 624117, 624119, 624121, 624151, 624158, 624162, 624167, 624170, 624618, 624622, 624643, 624678, 624679, 624682, 624683, 624720, 622539, 622540, 622570, 622572, 623734, 623748, 624739, and 624741.

Distribution pattern

US Nationwide Distribution including these states: NE, IL, FL, TX, OH, PA, IN, CA, NY, AL, ND, GA, AK, MI, WA, MS, NJ, IA, AZ, and RI.

Field note

Send feedback

We'll only use this to respond to your feedback.