VITROS 3600 Chemistry System, Catalog Number 6802783, Unique Device Identifier Number 10758750009930, Software Product Code 6802866, Software Version 3.2 and below; IVD. Product Usage: For use in the in vitro quantitative, semi-quantitative, and qualitative measurement of a variety of analytes of clinical interest, using VITROS Immunodiagnostic Products Reagents.
- Recall number
- Z-0640-2016
- Initiated
- November 03, 2015
- Classification
- Class II
- Status
- Terminated
- Recalling firm
- Ortho-Clinical Diagnostics
- Quantity
- Domestic: 200, Foreign: 554
App-derived interpretation
Software anomaly
Official device-enrichment evidence · Sourced
Software design
Inspect official wording and provenance
Reason for recall
Software anomaly may allow VITROS Systems to sample and process assays while the fluids and reagent temperatures are not in the required temperature range for optimal processing. When this intermittent issue occurs, the VITROS Systems continue to operate and process results without notification to the User, possibly producing erroneous results.
Code information
Software Version 3.2 and below; Serial Numbers J36000012 - 36000879
Distribution pattern
Worldwide Distribution - US Nationwide and the countries of Puerto Rico, and foreign distribution to: Argentina, Australia, Brazil, Canada, Chile, China, Colombia, England, France, Germany, India, Israel, Italy, Japan, Kenya, Mexico, Panama, Philippines, Poland, Romania, Russia, Singapore, Spain, Taiwan, Turkey, and Venezuela.