device · product 1 of 1
Ceiling Mounted Surgical Light System TruLight 3000/5000 and Ceiling Mounted Surgical Light System iLED3/iLED5
- Recall number
- Z-0563-2016
- Initiated
- November 02, 2015
- Classification
- Class II
- Status
- Completed
- Recalling firm
- Trumpf Medical Systems, Inc.
- Quantity
- 20,713
App-derived interpretation
The firm has received 15 complaints over ten years related to the connection of the spring arm to the top horizontal arm of the lighting system that while positioning the light head enabled the spring arm and light head to fall.
Official device-enrichment evidence · Sourced
Other
Inspect official wording and provenance
Reason for recall
The firm has received 15 complaints over ten years related to the connection of the spring arm to the top horizontal arm of the lighting system that while positioning the light head enabled the spring arm and light head to fall.
Code information
4028110 - iLED Single; 4028210 - iLED duo; 4028310 - iLED trio/quad; 4038110 - TruLight single; 4038210 - TruLight duo; 4038310 - TruLight trio/quad; 1565068 - iLED single (TUSM); 1565160 - iLED duo (TUSM); 1565161 - iLED trio/quad (TUSM); 1574759 - TruLight single (TUSM); 1574850 - TruLight duo (TUSM); 1574851 - TruLight trio/quad (TUSM)
Distribution pattern
USA (nationwide) Distribution to the states of : AL, AK, AZ, CA, CO, CT, DE, FL, GA, ID, IL, IN, IA, KY, ME, MA, MN, MO, NE, NV, NJ, NM, NY, NC, ND, OH, OR, PA, SC, TN, TX, UT, VA, VT, WA, WI and WY., and Internationally to Canada.