Skip to content
Recall Observatory FDA recall evidence

Exact FDA event identifier

Event 71983

1 recalled-product record grouped only because every row carries this exact official event ID.

Open FDA event record

Event summary

Timeline bucket
August 03, 2015
Product types
Device
Classifications
Class II
Statuses
Terminated
Recalling firm wording
Ferno-Washington Inc

Dossier provenance

Source snapshots represented here

  • openFDA Device Enforcement Report-date coverage June 20, 2012–July 08, 2026 Known gap: openFDA describes 2004-present coverage, but the current export has no report-date records in this interval.
  • openFDA Device Recall event_date_initiated coverage June 01, 1997–June 30, 2026

Enforcement coverage uses report dates; product initiation dates can precede those bounds. Device-enrichment coverage uses FDA event_date_initiated.

How coverage and grouping work

Complete imported group

Every recalled product in this event

1 official enforcement rows

Product dates, classifications, firms, and source wording remain attached to their individual rows. Opening this dossier never hides products that did not match a prior timeline filter.

device · product 1 of 1

Ferno iNX Wheeled Stretcher, Integrated Patient Transport & Loading System Product Usage: Stretcher for pre-hospital patient transport.

Z-2826-2015
Recall number
Z-2826-2015
Initiated
August 03, 2015
Classification
Class II
Status
Terminated
Recalling firm
Ferno-Washington Inc
Quantity
72 units

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
Hole elongation in the telescoping load frame channel. A hole elongation could affect the ability of the retention pin to stay seated which is necessary to preclude movement of the telescoping load frame.

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Sourced

Device Design

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

Hole elongation in the telescoping load frame channel. A hole elongation could affect the ability of the retention pin to stay seated which is necessary to preclude movement of the telescoping load frame.

Code information

Serial #s: 15F001703, 15F001704, 15F001706, 15F001707, 15F001708, 15F001709, 15F001710, 15F001711, 15F001712, 15F001713, 15F001715, 15F001717, 15F001719, 15F001721, 15F001722, 15F001723, 15F001724, 15F001725, 15F001726, 15F001727, 15F001728, 15F001729, 15F001730, 15F001731, 15F001732, 15F001733, 15F001734, 15F001735, 15F001736, 15F001737, 15F001738, 15F001739, 15F001740, 15F001741, 15F001742, 15F001744, 15F001745, 15F001746, 15F001747, 15F001748, 15F001749, 15F001750, 15F001751, 15F001752, 15F001753, 15F001754, 15F001755, 15F001756, 15F001757, 15F001758, 15F001759, 15F001760, 15F001761, 15F001762, 15F001763, 15F001764, 15F001765, 15F001766, 15F001767, 15F001768, 15F001769, , 5F001770, 15F001771, 15F001772, 15F001773, 15F001774, 15F001775, 15F001776, 15F001777, 15F001778, 15F001779, 15F001780, 15F001781, 15F001782, 15F001783, 15F001784, 15F001785, 15F001786, 15F001787, 15F001788, 15F001789, 15F001790, 15F001791, 15F001793, 15F001794, 15F001795, 15F001796, 15F001797, 15F001798, 15F001804, 15F001805, 15F001806, 15F001807, 15F001808, 15F001809, 15F001810, 15F001811, 15F001812, 15F001813, 15F001814, 15F001815, 15F001816 & 15F001850.

Distribution pattern

Worldwide Distribution - US Nationwide in the states of: CO, IA, IL, IN, KY, MD, MN, MO, NC, NJ, NY, OH, OK, OR, TX, VA, WA & WY and the countries of CANADA, AUSTRALIA, UNITED KINGDOM & NEW ZEALAND.

Field note

Send feedback

We'll only use this to respond to your feedback.