device · product 1 of 1
Panoramic X-ray Model PC-1000 The PC-1000 will enable the user to take panoramic x-ray images. The PC-1000/Laser 1000 will enable the user to take panoramic x-ray images as well as cephalometric x-ray images.
- Recall number
- Z-2538-2015
- Initiated
- June 04, 2015
- Classification
- Class II
- Status
- Terminated
- Recalling firm
- Panoramic Rental Corp.
- Quantity
- 7885
App-derived interpretation
Jackscrew nut part failure potentially leading to device malfunction and patient/operator injury.
Official device-enrichment evidence · Sourced
Equipment maintenance
Inspect official wording and provenance
Reason for recall
Jackscrew nut part failure potentially leading to device malfunction and patient/operator injury.
Code information
Model Number: 800724-1, 800724-3; Serial Numbers 6001 to 13885, Manufactured Between 1996 and 2003
Distribution pattern
Worldwide distribution: US (nationwide) all 50 states, including Puerto Rico; and countries of Antigua, Armenia, Aruba, Austria, Canada, China, Colombia, Ecuador, Egypt, England, France, Germany, Greece, Haiti, Honduras, Indonesia, Japan, Kenya, Lithuania, Mexico, Panama, Saudi Arabia, Singapore, Taiwan, Turkey, UAE, Uganda, Venezuela, Vietnam, and Virgin Islands.