Event 71101

3 recalled-product records grouped only because every row carries this exact official event ID.

Open FDA event record

Event summary

Timeline bucket
April 20, 2015
Product types
Device
Classifications
Class II
Statuses
Terminated
Recalling firm wording
Siemens Medical Solutions USA, Inc

Dossier provenance

Source snapshots represented here

  • openFDA Device Enforcement Report-date coverage June 20, 2012–July 08, 2026 Known gap: openFDA describes 2004-present coverage, but the current export has no report-date records in this interval.
  • openFDA Device Recall event_date_initiated coverage June 01, 1997–June 30, 2026

Enforcement coverage uses report dates; product initiation dates can precede those bounds. Device-enrichment coverage uses FDA event_date_initiated.

How coverage and grouping work

Complete imported group

Every recalled product in this event

3 official enforcement rows

Product dates, classifications, firms, and source wording remain attached to their individual rows. Opening this dossier never hides products that did not match a prior timeline filter.

device · product 1 of 3

SOMATOM Emotion 6; the intended use of computed tomography is to produce cross-sectional images of the body by computer reconstruction of x-ray transmission data from either the same axial plane taken at different angels or spiral planes taken at different angles.

Z-1644-2015
Recall number
Z-1644-2015
Initiated
April 20, 2015
Classification
Class II
Status
Terminated
Quantity
75

App-derived interpretation

Device software or design reason.device_software_design · v1.0.0
Software bugs in VC20B SP0a or SP1 software versions may cause issues that could

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Sourced

Software design

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

Software bugs in VC20B SP0a or SP1 software versions may cause issues that could make it necessary to rescan patients. Syngo Main UI may crash if the Patient Browser is scrolled with arrow keys of the keyboard. Tomo images may freeze on the screen when zoomed in or out under the mode of CAREVision. Sporadic displaying error in WorkStream4D application. DB may lock under heavy and multitasks.

Code information

SOMATOM Emotion 6 (2007) - 10165888 with serial numbers: 29338 27232 27230 27231

Distribution pattern

Nationwide Distribution including Puerto Rico.

device · product 2 of 3

SOMATOM Emotion 16, the intended use of computed tomography is to produce cross-sectional images of the body by computer reconstruction of x-ray transmission data from either the same axial plane taken at different angels or spiral planes taken at different angles.

Z-1645-2015
Recall number
Z-1645-2015
Initiated
April 20, 2015
Classification
Class II
Status
Terminated
Quantity
75

App-derived interpretation

Device software or design reason.device_software_design · v1.0.0
Software bugs in VC20B SP0a or SP1 software versions may cause issues that could

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Sourced

Software design

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

Software bugs in VC20B SP0a or SP1 software versions may cause issues that could make it necessary to rescan patients. Syngo Main UI may crash if the Patient Browser is scrolled with arrow keys of the keyboard. Tomo images may freeze on the screen when zoomed in or out under the mode of CAREVision. Sporadic displaying error in WorkStream4D application. DB may lock under heavy and multitasks.

Code information

SOMATOM Emotion 16 (2007) 10165977 with serial numbers: 86514 32490 80553 32552 32530 86406 86558 80542 80461 86405 32512 80441 86431 80467 86425 80442 86429 86557 80434 32581 32637 32597 80448 80431 32624 32469 80446 32578 80430

Distribution pattern

Nationwide Distribution including Puerto Rico.

device · product 3 of 3

SOMATOM Perspective; the intended use of computed tomography is to produce cross-sectional images of the body by computer reconstruction of x-ray transmission data from either the same axial plane taken at different angels or spiral planes taken at different angles.

Z-1646-2015
Recall number
Z-1646-2015
Initiated
April 20, 2015
Classification
Class II
Status
Terminated
Quantity
75

App-derived interpretation

Device software or design reason.device_software_design · v1.0.0
Software bugs in VC20B SP0a or SP1 software versions may cause issues that could

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Sourced

Software design

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

Software bugs in VC20B SP0a or SP1 software versions may cause issues that could make it necessary to rescan patients. Syngo Main UI may crash if the Patient Browser is scrolled with arrow keys of the keyboard. Tomo images may freeze on the screen when zoomed in or out under the mode of CAREVision. Sporadic displaying error in WorkStream4D application. DB may lock under heavy and multitasks.

Code information

SOMATOM Perspective - 10495568 with serial numbers: 59019 59195 59060 59043 59025 59029 59189 59204 76954 59010 59164 59185 59048 59169 77107 77110 59011 59030 59171 59168 59017 59167 59162 59190 59047 59061 77121 59028 59165 59020 59181 59037 59035 59005 59183 59016 59027 59184 76950 59203 59015 77182

Distribution pattern

Nationwide Distribution including Puerto Rico.