device · product 1 of 1
Polaris 4.75 TI Spinal System, 4.0 x 20 MM, Closed MAS, Closed Head Multiaxial Screw, Non-Sterile, Rx Only, REF 14-583020, Biomet Spine and Bone Healing Technologies. Spinal fixation component.
- Recall number
- Z-2152-2015
- Initiated
- March 12, 2015
- Classification
- Class II
- Status
- Terminated
- Recalling firm
- Biomet Spine, LLC
- Quantity
- 4,765 units
App-derived interpretation
out of specification
Official device-enrichment evidence · Sourced
Nonconforming Material/Component
Inspect official wording and provenance
Reason for recall
Multiaxial screws may not meet internal requirements related to fatigue strength. Compression testing result and surface treatment are out of specification.
Code information
Part Numbers: 14-582020, 14-582022, 14-582025, 14-582030, 14-582035, 14-582040, 14-582120, 14-582122, 14-582125, 14-582127, 14-582130, 14-582135, 14-582140, 14-582220, 14-582222, 14-582225, 14-582230, 14-582235, 14-582240, 14-582245, 14-582250, 14-592320, 14-582325, 14-582330, 14-582335, 14-582340, 14-582345, 14-582350, 14-582530, 14-582535, 14-582540, 14-582545, 14-582550, 14-582565, 14-582570, 14-582575, 14-583020, 14-583025, 14-583030, 14-583035, 14-583040, 14-583120, 14-583125, 14-583130, 14-583135, 14-583140, 14-583220, 14-583225, 14-583230, 14-583235, 14-583240, 14-583245, 14-583250, 14-583320, 14-583325, 14-583330, 14-583335, 14-583340, 14-583345 and 14-583350. All lots.
Distribution pattern
Distribution US Nationwide (AL, CA, CO, FL, GA, LA, MA, MD, MI, MO, NY, OH, PA, TX, WA, WI) and The Netherlands.