Skip to content
Recall Observatory FDA recall evidence

Exact FDA event identifier

Event 70472

1 recalled-product record grouped only because every row carries this exact official event ID.

Open FDA event record

Event summary

Timeline bucket
February 05, 2015
Product types
Device
Classifications
Class II
Statuses
Terminated
Recalling firm wording
Medical Components, Inc dba MedComp

Dossier provenance

Source snapshots represented here

  • openFDA Device Enforcement Report-date coverage June 20, 2012–July 08, 2026 Known gap: openFDA describes 2004-present coverage, but the current export has no report-date records in this interval.
  • openFDA Device Recall event_date_initiated coverage June 01, 1997–June 30, 2026

Enforcement coverage uses report dates; product initiation dates can precede those bounds. Device-enrichment coverage uses FDA event_date_initiated.

How coverage and grouping work

Complete imported group

Every recalled product in this event

1 official enforcement rows

Product dates, classifications, firms, and source wording remain attached to their individual rows. Opening this dossier never hides products that did not match a prior timeline filter.

device · product 1 of 1

Medcomp Duo-Flow 400XL Catheter, Hemo-Flow Catheter, SLX Catheter, Triple Lumen Infusion Catheter, CT rated PICC, Split Cath III, PC Split Cath III; AngioDynamics product- Dura-Flow Catheter, Schon XL Catheter; The affected product are insertion kits used to implant the devices packaged-long and short term hemodialysis catheters and infusion catheters.

Z-1351-2015
Recall number
Z-1351-2015
Initiated
February 05, 2015
Classification
Class II
Status
Terminated
Quantity
3152

App-derived interpretation

Potency or specification failure reason.potency_specification_failure · v1.0.0
out of specification

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Sourced

Process control

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

Medcomp has initiated the recall of Duo-Flow 400XL Catheter, Hemo-Flow Catheter, SLX Catheter, Triple Lumen Infusion Catheter, CT rated PICC, Split Cath III, PC Split Cath III because the product shipped was out of specification. One BI (biological indicator) out of 21 on the sterilization load was out of specification.

Code information

Medcomp products:Catalog Code/ Lot Number/Expiration Date: DFXL148MTE/ MGMC610/10/31/2019; HFS 32/ MGMB440/ 07/14/2017; HFS24E./MGLY830/07/31/2017; MC061402/MGMD190/07/15/2017; MC3L-8S/MGMD920/10/16/2019; MR17035211/MGMD680/10/31/2017; TRAY #593-3/MGMB460/07/15/2017; RMS23602/MBWM930/07/31/2017; ASPC2816-3/MBWM900/07/31/2017; Tray #55216-3/MBWL780/07/31/2017; ASPC2816-3PC/MBWX290/07/31/2017; 10301207/MGMD930/07/31/2017; 10800701/ MGLY730/ 07/31/2017; 10800702/ MGMD210/ 07/31/2017; 10800703/MGLY740/07/31/2017 AngioDynamics products- model numbers/lot numbers/ expiration dates: 10301207/MGMD930/07/31/2017; 10800701/MGLY730/07/31/2017; 10800702/MGMD210/07/31/2017; 10800703/MGLY740/07/31/2017

Distribution pattern

Worldwide Distribution.

Field note

Send feedback