device · product 1 of 1
Precise Digital Accelerator Intended to be used for radiation therapy treatments of malignant neoplastic diseases, as determined by a licensed medical practitioner.
- Recall number
- Z-1088-2015
- Initiated
- January 07, 2015
- Classification
- Class II
- Status
- Terminated
- Recalling firm
- Elekta, Inc.
- Quantity
- 46
App-derived interpretation
Elekta has identified a batch of securing bolts that fix the Gantry counter frame weight stack may fail.
Official device-enrichment evidence · Sourced
Other
Inspect official wording and provenance
Reason for recall
Elekta has identified a batch of securing bolts that fix the Gantry counter frame weight stack may fail.
Code information
153602, 153538, 153561, 153569, 153532, 153556, 153566, 153563, 153564, 153589, 153530, 153531, 153562, 153584, 153615, 153534, 153534, 153535, 153563, 153542, 153542, 153548, 153548, 153558, 153578, 153595, 153699, 153630, 153728, 153680, 153712. 153717, 153723, 153734, 153702, 153715, 153684, 153710, 153577, 153725, 153726, 153691, 153696, 153713, 153724, 153727
Distribution pattern
Worldwide Distribution - USA including AZ, CA, FL, KS, LA, MN, MO, NY, NC and Internationally to Austria, Bulgaria, Canada, China, Germany, Hungary, Italy, Malaysia, Malta, Netherlands, Norway, Russia, Spain, Sweden, Taiwan and Ukraine.