Skip to content
Recall Observatory FDA recall evidence

Exact FDA event identifier

Event 70086

1 recalled-product record grouped only because every row carries this exact official event ID.

Open FDA event record

Event summary

Timeline bucket
December 17, 2014
Product types
Device
Classifications
Class II
Statuses
Terminated
Recalling firm wording
Aesculap, Inc.

Dossier provenance

Source snapshots represented here

  • openFDA Device Enforcement Report-date coverage June 20, 2012–July 08, 2026 Known gap: openFDA describes 2004-present coverage, but the current export has no report-date records in this interval.
  • openFDA Device Recall event_date_initiated coverage June 01, 1997–June 30, 2026

Enforcement coverage uses report dates; product initiation dates can precede those bounds. Device-enrichment coverage uses FDA event_date_initiated.

How coverage and grouping work

Complete imported group

Every recalled product in this event

1 official enforcement rows

Product dates, classifications, firms, and source wording remain attached to their individual rows. Opening this dossier never hides products that did not match a prior timeline filter.

device · product 1 of 1

Miethke Shunt System accessories Product Usage: - The Miethke Shunt System is intended to shut cerebrospinal fluid from the lateral ventricles of the brain into the peritoneum.

Z-1012-2015
Recall number
Z-1012-2015
Initiated
December 17, 2014
Classification
Class II
Status
Terminated
Recalling firm
Aesculap, Inc.
Quantity
772

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
Some Miethke Shunt System accessories may have an incorrect MR safety symbol on the shelf box label. The label should have NO symbol but instead is marked with the MR conditional symbol. The product is MR unsafe. This information is only required to be in the Instructions for Use (IFU).

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Sourced

Labeling False and Misleading

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

Some Miethke Shunt System accessories may have an incorrect MR safety symbol on the shelf box label. The label should have NO symbol but instead is marked with the MR conditional symbol. The product is MR unsafe. This information is only required to be in the Instructions for Use (IFU).

Code information

material numbers: FV400T FV401T FV402T FV403T FV404T FV405T FV406T FV407T FV408T FV409T FV499T FV790T FV791T FV792T FV793T FV794T FV795T FV796T

Distribution pattern

Worldwide Distribution - US Nationwide

Field note

Send feedback

We'll only use this to respond to your feedback.