Skip to content
Recall Observatory FDA recall evidence

Exact FDA event identifier

Event 69954

2 recalled-product records grouped only because every row carries this exact official event ID.

Open FDA event record

Event summary

Timeline bucket
November 21, 2014
Product types
Device
Classifications
Class II
Statuses
Terminated
Recalling firm wording
Becton Dickinson & Co.

Dossier provenance

Source snapshots represented here

  • openFDA Device Enforcement Report-date coverage June 20, 2012–July 08, 2026 Known gap: openFDA describes 2004-present coverage, but the current export has no report-date records in this interval.
  • openFDA Device Recall event_date_initiated coverage June 01, 1997–June 30, 2026

Enforcement coverage uses report dates; product initiation dates can precede those bounds. Device-enrichment coverage uses FDA event_date_initiated.

How coverage and grouping work

Complete imported group

Every recalled product in this event

2 official enforcement rows

Product dates, classifications, firms, and source wording remain attached to their individual rows. Opening this dossier never hides products that did not match a prior timeline filter.

device · product 1 of 2

BBL Port A Cul tubes, catalog number 221606. Patient sample collection tubes. BBL Port A Cul tubes, jars and vials contain a reduced transport medium and are intended to maintain the viability of anaerobic, facultative and aerobic microorganisms during transport from the patient to the laboratory. Sterile packages are for collection of specimens in clean areas; e.g., surgical suites.

Z-1746-2015
Recall number
Z-1746-2015
Initiated
November 21, 2014
Classification
Class II
Status
Terminated
Recalling firm
Becton Dickinson & Co.
Quantity
145,860 tubes

App-derived interpretation

Microbial contamination reason.microbial_contamination · v1.0.0
Excessive oxidation may affect the viability of anaerobic microorganisms

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Sourced

Process control

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

The products may exhibit indications of excessive oxidation. Excessive oxidation may affect the viability of anaerobic microorganisms.

Code information

Part/Cat No. Lot / Serial No.Date of Mfr Exp. Date 221606 4042455 2/25/2014 2/5/2015 221606 4067180 3/18/2014 3/5/2015 221606 4087204 4/10/2014 3/25/2015 221606 4124702 5/8/2014 4/24/2015 221606 4161968 7/7/2014 6/12/2015 221606 4181959 7/14/2014 7/3/2015 221606 4032407 2/14/2014 1/29/2015 221606 4042069 2/17/2014 2/4/2015

Distribution pattern

US Nationwide distribution in AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NJ, NV, NY, OH, OR, PA, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV, and WY.

device · product 2 of 2

BBL Port A Cul tubes with Swabs Sterile Pack, catalog number 221607. For patient sample collection.

Z-1747-2015
Recall number
Z-1747-2015
Initiated
November 21, 2014
Classification
Class II
Status
Terminated
Recalling firm
Becton Dickinson & Co.
Quantity
92,710 units

App-derived interpretation

Microbial contamination reason.microbial_contamination · v1.0.0
Excessive oxidation may affect the viability of anaerobic microorganisms

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Sourced

Process control

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

The products may exhibit indications of excessive oxidation. Excessive oxidation may affect the viability of anaerobic microorganisms.

Code information

Part/Cat No. Lot / Serial No.Date of Mfr Exp. Date 221607 4067181 3/19/2014 3/5/2015 221607 4083274 4/3/2014 3/20/2015 221607 4124706 5/9/2014 4/24/2015 221607 4127564 5/22/2014 5/1/2015 221607 4170891 7/8/2014 6/19/2015 221607 4006235 2/18/2014 12/26/2014 221607 4028420 2/19/2014 1/21/2015

Distribution pattern

US Nationwide distribution in AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NJ, NV, NY, OH, OR, PA, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV, and WY.

Field note