Event 69927

12 recalled-product records grouped only because every row carries this exact official event ID.

Open FDA event record

Event summary

Timeline bucket
December 03, 2014
Product types
Device
Classifications
Class II
Statuses
Terminated
Recalling firm wording
Trumpf Medical Systems, Inc.

Dossier provenance

Source snapshots represented here

  • openFDA Device Enforcement Report-date coverage June 20, 2012–July 08, 2026 Known gap: openFDA describes 2004-present coverage, but the current export has no report-date records in this interval.
  • openFDA Device Recall event_date_initiated coverage June 01, 1997–June 30, 2026

Enforcement coverage uses report dates; product initiation dates can precede those bounds. Device-enrichment coverage uses FDA event_date_initiated.

How coverage and grouping work

Complete imported group

Every recalled product in this event

12 official enforcement rows

Product dates, classifications, firms, and source wording remain attached to their individual rows. Opening this dossier never hides products that did not match a prior timeline filter.

device · product 1 of 12

Cable remote control JUPITER U, Material no.4500119, Serial number: to (incl.) 102306915. Patient positioning during surgery, ranging from anesthesia induction to actual surgery to recovery from anesthesia. Patient transport on the operating tabletop, from the patient transfer system to the operating theatre or from the operating theatre to the patient transfer system.

Z-0918-2015
Recall number
Z-0918-2015
Initiated
December 03, 2014
Classification
Class II
Status
Terminated
Quantity
137 units

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
Complaints have been reported of unintended movement occurring during surgical cases due to the use of damaged JUPITER remotes

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Sourced

Component change control

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

Complaints have been reported of unintended movement occurring during surgical cases due to the use of damaged JUPITER remotes

Code information

The operating tables which are operated with the affected remote controls (all serial numbers): Mobile column JUPITER System U, Material no. 4500113; Mobile column JUPITER System Miyabi U, Material no. 1574977; Operating table column JUPITER SM 360 U, Material no. 1389824; Operating table column JUPITER SF U OCE, Material no. 1622774; Stationary column JUPITER System U, Material no. 4500112; FLOOR MOUNTING COLUMN JUPITER U, Material no. 1276277; Floor mounting column JUPITER 360 U, Material no.1389826; Operating table JUPITER UNIVERSAL U, Material no. 4500170; OR-TABLE JUPITER UNIVERSAL CARBON U, Material no.1223188.

Distribution pattern

Worldwide Distribution - USA including AL, AZ, CA, CT, FL, GA, IL, KS, LA, MD, MI, MK, MS, NC, NH, NJ, NY, OH, OK, PA, RI, SC, TX, VA, WA, WI and Internationally to Canada.

device · product 2 of 12

Cable remote control JUPITER / TS7500, Material no.1601926, Serial number: to (incl.) 101465919. Patient positioning during surgery, ranging from anesthesia induction to actual surgery to recovery from anesthesia. Patient transport on the operating tabletop, from the patient transfer system to the operating theatre or from the operating theatre to the patient transfer system.

Z-0919-2015
Recall number
Z-0919-2015
Initiated
December 03, 2014
Classification
Class II
Status
Terminated
Quantity
3303 total all devices

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
Complaints have been reported of unintended movement occurring during surgical cases due to the use of damaged JUPITER remotes

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Sourced

Component change control

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

Complaints have been reported of unintended movement occurring during surgical cases due to the use of damaged JUPITER remotes

Code information

The operating tables which are operated with the affected remote controls (all serial numbers): Mobile column JUPITER System U, Material no. 4500113; Mobile column JUPITER System Miyabi U, Material no. 1574977; Operating table column JUPITER SM 360 U, Material no. 1389824; Operating table column JUPITER SF U OCE, Material no. 1622774; Stationary column JUPITER System U, Material no. 4500112; FLOOR MOUNTING COLUMN JUPITER U, Material no. 1276277; Floor mounting column JUPITER 360 U, Material no.1389826; Operating table JUPITER UNIVERSAL U, Material no. 4500170; OR-TABLE JUPITER UNIVERSAL CARBON U, Material no.1223188.

Distribution pattern

Worldwide Distribution - USA including AL, AZ, CA, CT, FL, GA, IL, KS, LA, MD, MI, MK, MS, NC, NH, NJ, NY, OH, OK, PA, RI, SC, TX, VA, WA, WI and Internationally to Canada.

device · product 3 of 12

Cable remote control JUPITER, Material no. 4544119, Serial number: to (incl.) 102284378. Patient positioning during surgery, ranging from anesthesia induction to actual surgery to recovery from anesthesia. Patient transport on the operating tabletop, from the patient transfer system to the operating theatre or from the operating theatre to the patient transfer system.

Z-0920-2015
Recall number
Z-0920-2015
Initiated
December 03, 2014
Classification
Class II
Status
Terminated
Quantity
3303 total all devices

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
Complaints have been reported of unintended movement occurring during surgical cases due to the use of damaged JUPITER remotes

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Sourced

Component change control

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

Complaints have been reported of unintended movement occurring during surgical cases due to the use of damaged JUPITER remotes

Code information

The operating tables which are operated with the affected remote controls (all serial numbers): Mobile column JUPITER System U, Material no. 4500113; Mobile column JUPITER System Miyabi U, Material no. 1574977; Operating table column JUPITER SM 360 U, Material no. 1389824; Operating table column JUPITER SF U OCE, Material no. 1622774; Stationary column JUPITER System U, Material no. 4500112; FLOOR MOUNTING COLUMN JUPITER U, Material no. 1276277; Floor mounting column JUPITER 360 U, Material no.1389826; Operating table JUPITER UNIVERSAL U, Material no. 4500170; OR-TABLE JUPITER UNIVERSAL CARBON U, Material no.1223188.

Distribution pattern

Worldwide Distribution - USA including AL, AZ, CA, CT, FL, GA, IL, KS, LA, MD, MI, MK, MS, NC, NH, NJ, NY, OH, OK, PA, RI, SC, TX, VA, WA, WI and Internationally to Canada.

device · product 4 of 12

IR remote control JUPITER U, Material no. 4500120, Serial number: to (incl.) 102306940. Patient positioning during surgery, ranging from anesthesia induction to actual surgery to recovery from anesthesia. Patient transport on the operating tabletop, from the patient transfer system to the operating theatre or from the operating theatre to the patient transfer system.

Z-0921-2015
Recall number
Z-0921-2015
Initiated
December 03, 2014
Classification
Class II
Status
Terminated
Quantity
3303 total all devices

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
Complaints have been reported of unintended movement occurring during surgical cases due to the use of damaged JUPITER remotes

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Sourced

Component change control

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

Complaints have been reported of unintended movement occurring during surgical cases due to the use of damaged JUPITER remotes

Code information

The operating tables which are operated with the affected remote controls (all serial numbers): Mobile column JUPITER System U, Material no. 4500113; Mobile column JUPITER System Miyabi U, Material no. 1574977; Operating table column JUPITER SM 360 U, Material no. 1389824; Operating table column JUPITER SF U OCE, Material no. 1622774; Stationary column JUPITER System U, Material no. 4500112; FLOOR MOUNTING COLUMN JUPITER U, Material no. 1276277; Floor mounting column JUPITER 360 U, Material no.1389826; Operating table JUPITER UNIVERSAL U, Material no. 4500170; OR-TABLE JUPITER UNIVERSAL CARBON U, Material no.1223188.

Distribution pattern

Worldwide Distribution - USA including AL, AZ, CA, CT, FL, GA, IL, KS, LA, MD, MI, MK, MS, NC, NH, NJ, NY, OH, OK, PA, RI, SC, TX, VA, WA, WI and Internationally to Canada.

device · product 5 of 12

IR remote control JUPITER / TS7500, Material no.1601927, Serial number: to (incl.) 102222333. Patient positioning during surgery, ranging from anesthesia induction to actual surgery to recovery from anesthesia. Patient transport on the operating tabletop, from the patient transfer system to the operating theatre or from the operating theatre to the patient transfer system.

Z-0922-2015
Recall number
Z-0922-2015
Initiated
December 03, 2014
Classification
Class II
Status
Terminated
Quantity
3303 total all devices

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
Complaints have been reported of unintended movement occurring during surgical cases due to the use of damaged JUPITER remotes

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Sourced

Component change control

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

Complaints have been reported of unintended movement occurring during surgical cases due to the use of damaged JUPITER remotes

Code information

The operating tables which are operated with the affected remote controls (all serial numbers): Mobile column JUPITER System U, Material no. 4500113; Mobile column JUPITER System Miyabi U, Material no. 1574977; Operating table column JUPITER SM 360 U, Material no. 1389824; Operating table column JUPITER SF U OCE, Material no. 1622774; Stationary column JUPITER System U, Material no. 4500112; FLOOR MOUNTING COLUMN JUPITER U, Material no. 1276277; Floor mounting column JUPITER 360 U, Material no.1389826; Operating table JUPITER UNIVERSAL U, Material no. 4500170; OR-TABLE JUPITER UNIVERSAL CARBON U, Material no.1223188.

Distribution pattern

Worldwide Distribution - USA including AL, AZ, CA, CT, FL, GA, IL, KS, LA, MD, MI, MK, MS, NC, NH, NJ, NY, OH, OK, PA, RI, SC, TX, VA, WA, WI and Internationally to Canada.

device · product 6 of 12

IIR remote control JUPITER, Material no.4544120, Serial number: to (incl.) 102006720. Patient positioning during surgery, ranging from anesthesia induction to actual surgery to recovery from anesthesia. Patient transport on the operating tabletop, from the patient transfer system to the operating theatre or from the operating theatre to the patient transfer system.

Z-0923-2015
Recall number
Z-0923-2015
Initiated
December 03, 2014
Classification
Class II
Status
Terminated
Quantity
3303 total all devices

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
Complaints have been reported of unintended movement occurring during surgical cases due to the use of damaged JUPITER remotes

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Sourced

Component change control

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

Complaints have been reported of unintended movement occurring during surgical cases due to the use of damaged JUPITER remotes

Code information

The operating tables which are operated with the affected remote controls (all serial numbers): Mobile column JUPITER System U, Material no. 4500113; Mobile column JUPITER System Miyabi U, Material no. 1574977; Operating table column JUPITER SM 360 U, Material no. 1389824; Operating table column JUPITER SF U OCE, Material no. 1622774; Stationary column JUPITER System U, Material no. 4500112; FLOOR MOUNTING COLUMN JUPITER U, Material no. 1276277; Floor mounting column JUPITER 360 U, Material no.1389826; Operating table JUPITER UNIVERSAL U, Material no. 4500170; OR-TABLE JUPITER UNIVERSAL CARBON U, Material no.1223188.

Distribution pattern

Worldwide Distribution - USA including AL, AZ, CA, CT, FL, GA, IL, KS, LA, MD, MI, MK, MS, NC, NH, NJ, NY, OH, OK, PA, RI, SC, TX, VA, WA, WI and Internationally to Canada.

device · product 7 of 12

(R)IR remote control JUPITER, Material no.1219656, Serial number: to (incl.) 102317869. Patient positioning during surgery, ranging from anesthesia induction to actual surgery to recovery from anesthesia. Patient transport on the operating tabletop, from the patient transfer system to the operating theatre or from the operating theatre to the patient transfer system.

Z-0924-2015
Recall number
Z-0924-2015
Initiated
December 03, 2014
Classification
Class II
Status
Terminated
Quantity
3303 total all devices

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
Complaints have been reported of unintended movement occurring during surgical cases due to the use of damaged JUPITER remotes

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Sourced

Component change control

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

Complaints have been reported of unintended movement occurring during surgical cases due to the use of damaged JUPITER remotes

Code information

The operating tables which are operated with the affected remote controls (all serial numbers): Mobile column JUPITER System U, Material no. 4500113; Mobile column JUPITER System Miyabi U, Material no. 1574977; Operating table column JUPITER SM 360 U, Material no. 1389824; Operating table column JUPITER SF U OCE, Material no. 1622774; Stationary column JUPITER System U, Material no. 4500112; FLOOR MOUNTING COLUMN JUPITER U, Material no. 1276277; Floor mounting column JUPITER 360 U, Material no.1389826; Operating table JUPITER UNIVERSAL U, Material no. 4500170; OR-TABLE JUPITER UNIVERSAL CARBON U, Material no.1223188.

Distribution pattern

Worldwide Distribution - USA including AL, AZ, CA, CT, FL, GA, IL, KS, LA, MD, MI, MK, MS, NC, NH, NJ, NY, OH, OK, PA, RI, SC, TX, VA, WA, WI and Internationally to Canada.

device · product 8 of 12

(R)Cable remote control JUPITER, Material no.1219655, Serial number: to (incl.) 101998119. Patient positioning during surgery, ranging from anesthesia induction to actual surgery to recovery from anesthesia. Patient transport on the operating tabletop, from the patient transfer system to the operating theatre or from the operating theatre to the patient transfer system.

Z-0925-2015
Recall number
Z-0925-2015
Initiated
December 03, 2014
Classification
Class II
Status
Terminated
Quantity
3303 total all devices

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
Complaints have been reported of unintended movement occurring during surgical cases due to the use of damaged JUPITER remotes

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Sourced

Component change control

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

Complaints have been reported of unintended movement occurring during surgical cases due to the use of damaged JUPITER remotes

Code information

The operating tables which are operated with the affected remote controls (all serial numbers): Mobile column JUPITER System U, Material no. 4500113; Mobile column JUPITER System Miyabi U, Material no. 1574977; Operating table column JUPITER SM 360 U, Material no. 1389824; Operating table column JUPITER SF U OCE, Material no. 1622774; Stationary column JUPITER System U, Material no. 4500112; FLOOR MOUNTING COLUMN JUPITER U, Material no. 1276277; Floor mounting column JUPITER 360 U, Material no.1389826; Operating table JUPITER UNIVERSAL U, Material no. 4500170; OR-TABLE JUPITER UNIVERSAL CARBON U, Material no.1223188.

Distribution pattern

Worldwide Distribution - USA including AL, AZ, CA, CT, FL, GA, IL, KS, LA, MD, MI, MK, MS, NC, NH, NJ, NY, OH, OK, PA, RI, SC, TX, VA, WA, WI and Internationally to Canada.

device · product 9 of 12

Cable remote control JUX Material, no.1307315, Serial number: to (incl.) 101287649. Patient positioning during surgery, ranging from anesthesia induction to actual surgery to recovery from anesthesia. Patient transport on the operating tabletop, from the patient transfer system to the operating theatre or from the operating theatre to the patient transfer system.

Z-0926-2015
Recall number
Z-0926-2015
Initiated
December 03, 2014
Classification
Class II
Status
Terminated
Quantity
3303 total all devices

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
Complaints have been reported of unintended movement occurring during surgical cases due to the use of damaged JUPITER remotes

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Sourced

Component change control

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

Complaints have been reported of unintended movement occurring during surgical cases due to the use of damaged JUPITER remotes

Code information

The operating tables which are operated with the affected remote controls (all serial numbers): Mobile column JUPITER System U, Material no. 4500113; Mobile column JUPITER System Miyabi U, Material no. 1574977; Operating table column JUPITER SM 360 U, Material no. 1389824; Operating table column JUPITER SF U OCE, Material no. 1622774; Stationary column JUPITER System U, Material no. 4500112; FLOOR MOUNTING COLUMN JUPITER U, Material no. 1276277; Floor mounting column JUPITER 360 U, Material no.1389826; Operating table JUPITER UNIVERSAL U, Material no. 4500170; OR-TABLE JUPITER UNIVERSAL CARBON U, Material no.1223188.

Distribution pattern

Worldwide Distribution - USA including AL, AZ, CA, CT, FL, GA, IL, KS, LA, MD, MI, MK, MS, NC, NH, NJ, NY, OH, OK, PA, RI, SC, TX, VA, WA, WI and Internationally to Canada.

device · product 10 of 12

IR remote control JUX, Material no.1307314, Serial number: to (incl.) 102338310 Patient positioning during surgery, ranging from anesthesia induction to actual surgery to recovery from anesthesia. Patient transport on the operating tabletop, from the patient transfer system to the operating theatre or from the operating theatre to the patient transfer system.

Z-0927-2015
Recall number
Z-0927-2015
Initiated
December 03, 2014
Classification
Class II
Status
Terminated
Quantity
3303 total all devices

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
Complaints have been reported of unintended movement occurring during surgical cases due to the use of damaged JUPITER remotes

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Sourced

Component change control

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

Complaints have been reported of unintended movement occurring during surgical cases due to the use of damaged JUPITER remotes

Code information

The operating tables which are operated with the affected remote controls (all serial numbers): Mobile column JUPITER System U, Material no. 4500113; Mobile column JUPITER System Miyabi U, Material no. 1574977; Operating table column JUPITER SM 360 U, Material no. 1389824; Operating table column JUPITER SF U OCE, Material no. 1622774; Stationary column JUPITER System U, Material no. 4500112; FLOOR MOUNTING COLUMN JUPITER U, Material no. 1276277; Floor mounting column JUPITER 360 U, Material no.1389826; Operating table JUPITER UNIVERSAL U, Material no. 4500170; OR-TABLE JUPITER UNIVERSAL CARBON U, Material no.1223188.

Distribution pattern

Worldwide Distribution - USA including AL, AZ, CA, CT, FL, GA, IL, KS, LA, MD, MI, MK, MS, NC, NH, NJ, NY, OH, OK, PA, RI, SC, TX, VA, WA, WI and Internationally to Canada.

device · product 11 of 12

Cable remote control JUC, BUTTON, Material no.1239823, Serial number: to (incl.) 100143048. Patient positioning during surgery, ranging from anesthesia induction to actual surgery to recovery from anesthesia. Patient transport on the operating tabletop, from the patient transfer system to the operating theatre or from the operating theatre to the patient transfer system.

Z-0928-2015
Recall number
Z-0928-2015
Initiated
December 03, 2014
Classification
Class II
Status
Terminated
Quantity
3303 total all devices

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
Complaints have been reported of unintended movement occurring during surgical cases due to the use of damaged JUPITER remotes

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Sourced

Component change control

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

Complaints have been reported of unintended movement occurring during surgical cases due to the use of damaged JUPITER remotes

Code information

The operating tables which are operated with the affected remote controls (all serial numbers): Mobile column JUPITER System U, Material no. 4500113; Mobile column JUPITER System Miyabi U, Material no. 1574977; Operating table column JUPITER SM 360 U, Material no. 1389824; Operating table column JUPITER SF U OCE, Material no. 1622774; Stationary column JUPITER System U, Material no. 4500112; FLOOR MOUNTING COLUMN JUPITER U, Material no. 1276277; Floor mounting column JUPITER 360 U, Material no.1389826; Operating table JUPITER UNIVERSAL U, Material no. 4500170; OR-TABLE JUPITER UNIVERSAL CARBON U, Material no.1223188.

Distribution pattern

Worldwide Distribution - USA including AL, AZ, CA, CT, FL, GA, IL, KS, LA, MD, MI, MK, MS, NC, NH, NJ, NY, OH, OK, PA, RI, SC, TX, VA, WA, WI and Internationally to Canada.

device · product 12 of 12

IR remote control JUC, BUTTON, Material no.1239824, Serial number: to (incl.) 100139854. Patient positioning during surgery, ranging from anesthesia induction to actual surgery to recovery from anesthesia. Patient transport on the operating tabletop, from the patient transfer system to the operating theatre or from the operating theatre to the patient transfer system.

Z-0929-2015
Recall number
Z-0929-2015
Initiated
December 03, 2014
Classification
Class II
Status
Terminated
Quantity
3303 total all devices

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
Complaints have been reported of unintended movement occurring during surgical cases due to the use of damaged JUPITER remotes

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Sourced

Component change control

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

Complaints have been reported of unintended movement occurring during surgical cases due to the use of damaged JUPITER remotes

Code information

The operating tables which are operated with the affected remote controls (all serial numbers): Mobile column JUPITER System U, Material no. 4500113; Mobile column JUPITER System Miyabi U, Material no. 1574977; Operating table column JUPITER SM 360 U, Material no. 1389824; Operating table column JUPITER SF U OCE, Material no. 1622774; Stationary column JUPITER System U, Material no. 4500112; FLOOR MOUNTING COLUMN JUPITER U, Material no. 1276277; Floor mounting column JUPITER 360 U, Material no.1389826; Operating table JUPITER UNIVERSAL U, Material no. 4500170; OR-TABLE JUPITER UNIVERSAL CARBON U, Material no.1223188.

Distribution pattern

Worldwide Distribution - USA including AL, AZ, CA, CT, FL, GA, IL, KS, LA, MD, MI, MK, MS, NC, NH, NJ, NY, OH, OK, PA, RI, SC, TX, VA, WA, WI and Internationally to Canada.