October 16, 2014
Product types
Device
Classifications
Class III
Statuses
Terminated
Recalling firm wording
Biomerieux

Dossier provenance

Source snapshots represented here

  • openFDA Device Enforcement Report-date coverage June 20, 2012–July 08, 2026 Known gap: openFDA describes 2004-present coverage, but the current export has no report-date records in this interval.
  • openFDA Device Recall event_date_initiated coverage June 01, 1997–June 30, 2026

Enforcement coverage uses report dates; product initiation dates can precede those bounds. Device-enrichment coverage uses FDA event_date_initiated.

How coverage and grouping work

Complete imported group

Every recalled product in this event

5 official enforcement rows

Product dates, classifications, firms, and source wording remain attached to their individual rows. Opening this dossier never hides products that did not match a prior timeline filter.

device · product 1 of 5

Etest¿ Ceftaroline (CPT32) WW B30. Etest¿ (Antimicrobial Susceptibility Testing) is a quantitative technique for determining the antimicrobial susceptibility of Gram negative and Gram positive aerobic bacteria.

Z-0230-2015
Recall number
Z-0230-2015
Initiated
October 16, 2014
Classification
Class III
Status
Terminated
Recalling firm
Biomerieux
Quantity
652 devices

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
The products have a wrong expiration date on their labeling, 5 years instead of 2 years.

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Sourced

Labeling Change Control

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

The products have a wrong expiration date on their labeling, 5 years instead of 2 years.

Code information

Ref #537500, Lot #1003160540, Right Expiry date: 27-APR-2016

Distribution pattern

Worldwide Distribution-US (nationwide) including DC and the states of AL, AK, AR, AZ, CA, CO, DE, FL, GA, IA, IN, KY, MA, MD, MI, MO, NC, NE, NY, OH, OK, PA, SC, TN, TX, UT, VA, WA, WI, and the countries of Australia, Austria, China, Czech Republic, France, Germany, India, Italy, Japan, Netherlands, Poland, Portugal, Switzerland, Sweden, Turkey, United Kingdom, Egypt, Pakistan, Slovenia and Uruguay.

device · product 2 of 5

Etest¿ Ceftaroline (CPT32) WWB100. Etest¿ (Antimicrobial Susceptibility Testing) is a quantitative technique for determining the antimicrobial susceptibility of Gram negative and Gram positive aerobic bacteria.

Z-0231-2015
Recall number
Z-0231-2015
Initiated
October 16, 2014
Classification
Class III
Status
Terminated
Recalling firm
Biomerieux
Quantity
652 devices

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
The products have a wrong expiration date on their labeling, 5 years instead of 2 years.

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Sourced

Labeling Change Control

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

The products have a wrong expiration date on their labeling, 5 years instead of 2 years.

Code information

Ref #537508, Lot #1003160510, Right Expiry date: 27-APR-2016

Distribution pattern

Worldwide Distribution-US (nationwide) including DC and the states of AL, AK, AR, AZ, CA, CO, DE, FL, GA, IA, IN, KY, MA, MD, MI, MO, NC, NE, NY, OH, OK, PA, SC, TN, TX, UT, VA, WA, WI, and the countries of Australia, Austria, China, Czech Republic, France, Germany, India, Italy, Japan, Netherlands, Poland, Portugal, Switzerland, Sweden, Turkey, United Kingdom, Egypt, Pakistan, Slovenia and Uruguay.

device · product 3 of 5

Etest¿ Ceftaroline (CPT32) US B30. Etest¿ (Antimicrobial Susceptibility Testing) is a quantitative technique for determining the antimicrobial susceptibility of Gram negative and Gram positive aerobic bacteria.

Z-0232-2015
Recall number
Z-0232-2015
Initiated
October 16, 2014
Classification
Class III
Status
Terminated
Recalling firm
Biomerieux
Quantity
652 devices

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
The products have a wrong expiration date on their labeling, 5 years instead of 2 years.

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Sourced

Labeling Change Control

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

The products have a wrong expiration date on their labeling, 5 years instead of 2 years.

Code information

Ref #537540, Lot #1003202200, Right Expiry date: 27-APR-2016

Distribution pattern

Worldwide Distribution-US (nationwide) including DC and the states of AL, AK, AR, AZ, CA, CO, DE, FL, GA, IA, IN, KY, MA, MD, MI, MO, NC, NE, NY, OH, OK, PA, SC, TN, TX, UT, VA, WA, WI, and the countries of Australia, Austria, China, Czech Republic, France, Germany, India, Italy, Japan, Netherlands, Poland, Portugal, Switzerland, Sweden, Turkey, United Kingdom, Egypt, Pakistan, Slovenia and Uruguay.

device · product 4 of 5

Etest¿ Ceftaroline (CPT32) US B100. Etest¿ (Antimicrobial Susceptibility Testing) is a quantitative technique for determining the antimicrobial susceptibility of Gram negative and Gram positive aerobic bacteria.

Z-0233-2015
Recall number
Z-0233-2015
Initiated
October 16, 2014
Classification
Class III
Status
Terminated
Recalling firm
Biomerieux
Quantity
652 devices

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
The products have a wrong expiration date on their labeling, 5 years instead of 2 years.

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Sourced

Labeling Change Control

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

The products have a wrong expiration date on their labeling, 5 years instead of 2 years.

Code information

Ref #537548, Lot #1003202820, Right Expiry date: 27-APR-2016

Distribution pattern

Worldwide Distribution-US (nationwide) including DC and the states of AL, AK, AR, AZ, CA, CO, DE, FL, GA, IA, IN, KY, MA, MD, MI, MO, NC, NE, NY, OH, OK, PA, SC, TN, TX, UT, VA, WA, WI, and the countries of Australia, Austria, China, Czech Republic, France, Germany, India, Italy, Japan, Netherlands, Poland, Portugal, Switzerland, Sweden, Turkey, United Kingdom, Egypt, Pakistan, Slovenia and Uruguay.

device · product 5 of 5

Etest¿ Ceftaroline (CPT32) US B30. Etest¿ (Antimicrobial Susceptibility Testing) is a quantitative technique for determining the antimicrobial susceptibility of Gram negative and Gram positive aerobic bacteria.

Z-0234-2015
Recall number
Z-0234-2015
Initiated
October 16, 2014
Classification
Class III
Status
Terminated
Recalling firm
Biomerieux
Quantity
652 devices

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
The products have a wrong expiration date on their labeling, 5 years instead of 2 years.

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Sourced

Labeling Change Control

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

The products have a wrong expiration date on their labeling, 5 years instead of 2 years.

Code information

Ref #537540, Lot #1003301830, Right Expiry date: 30-APR-2016

Distribution pattern

Worldwide Distribution-US (nationwide) including DC and the states of AL, AK, AR, AZ, CA, CO, DE, FL, GA, IA, IN, KY, MA, MD, MI, MO, NC, NE, NY, OH, OK, PA, SC, TN, TX, UT, VA, WA, WI, and the countries of Australia, Austria, China, Czech Republic, France, Germany, India, Italy, Japan, Netherlands, Poland, Portugal, Switzerland, Sweden, Turkey, United Kingdom, Egypt, Pakistan, Slovenia and Uruguay.