Skip to content
Recall Observatory FDA recall evidence

Exact FDA event identifier

Event 69253

1 recalled-product record grouped only because every row carries this exact official event ID.

Open FDA event record

Event summary

Timeline bucket
October 17, 2013
Product types
Device
Classifications
Class II
Statuses
Terminated
Recalling firm wording
Baxter Healthcare Corp

Dossier provenance

Source snapshots represented here

  • openFDA Device Enforcement Report-date coverage June 20, 2012–July 08, 2026 Known gap: openFDA describes 2004-present coverage, but the current export has no report-date records in this interval.
  • openFDA Device Recall event_date_initiated coverage June 01, 1997–June 30, 2026

Enforcement coverage uses report dates; product initiation dates can precede those bounds. Device-enrichment coverage uses FDA event_date_initiated.

How coverage and grouping work

Complete imported group

Every recalled product in this event

1 official enforcement rows

Product dates, classifications, firms, and source wording remain attached to their individual rows. Opening this dossier never hides products that did not match a prior timeline filter.

device · product 1 of 1

15 L Cycler Drainage Bag Product Usage: For use with Baxter Cycler Tubing sets as a peritoneal dialysis accessory.

Z-0080-2015
Recall number
Z-0080-2015
Initiated
October 17, 2013
Classification
Class II
Status
Terminated
Recalling firm
Baxter Healthcare Corp
Quantity
3,085,080

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
The large tube clamp used to close the drain tube on the 15 L Cycler Drainage Bag may not fully close the drain tube and may result in leakage.

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Sourced

Component change control

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

The large tube clamp used to close the drain tube on the 15 L Cycler Drainage Bag may not fully close the drain tube and may result in leakage.

Code information

Lot numbers: H12K20044, H12K20085, H12K21026, H13B04052, H13B04086, H13B05042, H13B05067, H13B06073, H13B07063, H13B13087, H13B14069, H13B14085, H13B15066, H13B18029, H13B18084, H13B20108, H13B22013, H13B22062, H13B23060, H13B24043, H13B25032, H13B25107, H13C06030, H13C06055, H13C07053, H13C07087, H13C08051, H13C09026, H13C09034, H13C09075, H13C11089, H13C12046, H13C12061, H13C13069, H13C19124, H13C20031, H13C20106, H13C21054, H13C22029, H13C25048, H13C25071, H13C26038, H13C26137, H13D02086, H13D03068, H13D04033, H13D04108, H13D10089, H13D10097, H13D11087, H13D12051, H13D12101, H13D13034, H13D13067, H13D15070, H13D15112, H13D16078, H13D16110, H13D19056, H13D19098, H13D20013, H13D20062, H13D21029, H13D22043, H13D22084, H13D23025, H13D23041, H13D24064, H13D25020, H13D25095, H13E01036, H13E01077, H13E02026, H13E02091, H13E03016, H13E07116, H13E08023, H13E08049, H13E09070, H13E10011, H13E10029, H13E13015, H13E13031, H13E14088, H13E14104, H13E15036, H13E21083, H13E22073, H13E23022, H13E23055, H13E24012, H13E24061, H13E28047, H13E28088, H13E29029, H13E30035, H13E30092, H13F04095, H13F04129, H13F05084, H13F06025, H13F06060, H13F07056, H13F11033, H13F12049, H13F12072, H13F12122, H13F13039, H13F13070, H13F14094, H13F18087, H13F19036, H13F20018, H13F20091, H13F21016, H13F21073, H13F22014, H13F22063, H13F25041, H13F25090, H13F26031, H13F26072, H13F27013, H13F27054, H13F28029, H13G09084, H13G10025, H13G10124, H13G11064, H13G12039, H13G12088, H13G16105, H13G17046, H13G18010, H13G18101, H13G19067, H13G19109, H13G23069, H13G24018, H13G24075, H13G25015, H13G25023, H13G26013, H13G26070, H13G31054, H13G31104, H13H01055, H13H02012, H13H02095, H13H05049, H13H06054, H13H06120, H13H07045, H13H13019, H13H13035, H13H13092, H13H14025, H13H14090, H13H15048, H13H15105, H13H20030, H13H20063, H13H21046, H13H21079, H13H22028, H13H22077, H13H23042

Distribution pattern

Worldwide Distribution - US Nationwide (including Puerto Rico) and countries of, Canada, Australia, New Zealand, Thailand, Malaysia, Singapore, Korea, Japan, Philippines, Columbia, Ecuador, Austria, Bahran, Belgium, Bosnia & Herzegovina, Bulgaria, Czech Republic, Denmark, Estonia, Finland, France, Georgia, Germany, Great Britain, Greece, Hungary, Ireland, Isreal, Italy, Jordan, Kazakhstan, Latvia, Luxembourg, Malta, Norway, Oman, Poland, Portugal, Qatar, Russia, Saudi Arabia, Slovakia, South Africa, Spain (including Canary Islands), Sweden, Switzerland, The Netherlands, Turkey and United Arad Emirates.