Perouse Flamingo, Model 0218TA and 0218TB intended for use during cardiovascular procedure to create, maintain and monitor pressure in the balloon catheter.
- Recall number
- Z-2089-2014
- Initiated
- May 10, 2012
- Classification
- Class II
- Status
- Terminated
- Recalling firm
- Perouse Medical
- Quantity
- 2,496 units
App-derived interpretation
Perouse Medical in France initiated a voluntary recall of certain batches of Perouse FLAMINGO Inflations Devices due to a potential water-tightness default. The potential default could cause an air entry when the plunger is positioned around 20 mL.
Official device-enrichment evidence · Sourced
Process control
Inspect official wording and provenance
Reason for recall
Perouse Medical in France initiated a voluntary recall of certain batches of Perouse FLAMINGO Inflations Devices due to a potential water-tightness default. The potential default could cause an air entry when the plunger is positioned around 20 mL.
Code information
Model 0218TA and 0218TB. Manufacturing Lot Numbers: 12031275 and 12031242. EXP.: 02/2017. Batches: 12031526, 12031249, 12031527, 12031248, 12041183, 12041186, 12031244, 12031461, 12031460, 12021404, 12031462, 12031245, 12021405, 12031405, 12031246, 12021401, 12041344, 12031237, 12031241, 12021403, 12041345, 12041184, 12031497, 12031240, 12021194, 12031524, 12031480, 12041181, 12031275, 12031242, 12021450, 12021449, 12041185, 12031525, 12031247, 12021402, 12021448.
Distribution pattern
US.