Recall Observatory FDA recall evidence

Exact FDA event identifier

Event 68790

3 recalled-product records grouped only because every row carries this exact official event ID.

Open FDA event record

Event summary

Timeline bucket
July 07, 2014
Product types
Device
Classifications
Class II
Statuses
Terminated
Recalling firm wording
Ortho-Clinical Diagnostics

Dossier provenance

Source snapshots represented here

  • openFDA Device Enforcement Report-date coverage June 20, 2012–July 08, 2026 Known gap: openFDA describes 2004-present coverage, but the current export has no report-date records in this interval.
  • openFDA Device Recall event_date_initiated coverage June 01, 1997–June 30, 2026

Enforcement coverage uses report dates; product initiation dates can precede those bounds. Device-enrichment coverage uses FDA event_date_initiated.

How coverage and grouping work

Complete imported group

Every recalled product in this event

3 official enforcement rows

Product dates, classifications, firms, and source wording remain attached to their individual rows. Opening this dossier never hides products that did not match a prior timeline filter.

device · product 1 of 3

VITROS 3600 Immunodiagnostic System --- For use in the in vitro quantitative, semi-quantitative and qualitative measurement of a variety of analytes of clinical interest, using VITROS Immunodiagnostic Products Reagents. The VITROS Software is a part of the VITROS 3600, 4600, & 5600 Systems. The VITROS 3600, 4600, & 5600 Systems share the same software base. VITROS Sample Programming consists of Sample Identification (SID), test requests, and patient demographics. Patient demographics include patient name, address, date of birth, gender, physician, and patient identification (PID).

Z-2139-2014
Recall number
Z-2139-2014
Initiated
July 07, 2014
Classification
Class II
Status
Terminated
Quantity
Domestic: 142, Foreign: 536

App-derived interpretation

Device software or design reason.device_software_design · v1.0.0
anomaly with VITROS System Software

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Sourced

Software in the Use Environment

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

Ortho Clinical Diagnostics has identified an anomaly with VITROS System Software Version 3.0 and below. Internal testing confirmed that under very specific conditions, incorrect patient demographics were associated with a sample identification (ID) number.

Code information

Serial Numbers J36000101 - J36000806 (J Numbers are analogous to Serial Numbers and for VITROS 3600 systems the J Numbers start with 3600XXXX)

Distribution pattern

Worldwide Distribution-USA (nationwide) including Puerto Rico and the countries of Australia, Brazil, Canada, Chile, China, Colombia, England, France, Germany, India, Italy, Japan, Mexico, Panama, Singapore, Spain and Venezuela.

device · product 2 of 3

VITROS 4600 Chemistry System (VITROS 5,1 FS System family member) --- The VITROS 4600 Chemistry System is intended for use in the in vitro quantitative measurement of a variety of analytes of clinical interest, using both VITROS Chemistry Products Slides (colorimetric endpoint, rate, ion-selective electrode, and immunorate methods) and VITROS Chemistry Products MicroTip liquid reagents (spectrophotometric and spectrophotometric immunoassay methods.) The VITROS Software is a part of the VITROS 3600, 4600, & 5600 Systems. The VITROS 3600, 4600, & 5600 Systems share the same software base. VITROS Sample Programming consists of Sample Identification (SID), test requests, and patient demographics. Patient demographics include patient name, address, date of birth, gender, physician, and patient identification (PID).

Z-2140-2014
Recall number
Z-2140-2014
Initiated
July 07, 2014
Classification
Class II
Status
Terminated
Quantity
Domestic: 81, Foreign: 238

App-derived interpretation

Device software or design reason.device_software_design · v1.0.0
anomaly with VITROS System Software

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Sourced

Software in the Use Environment

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

Ortho Clinical Diagnostics has identified an anomaly with VITROS System Software Version 3.0 and below. Internal testing confirmed that under very specific conditions, incorrect patient demographics were associated with a sample identification (ID) number.

Code information

Serial Numbers J46000108 - J46000440 (J Numbers are analogous to Serial Numbers, for VITROS 4600 systems the J Numbers start with 4600XXXX)

Distribution pattern

Worldwide Distribution-USA (nationwide) including Puerto Rico and the countries of Australia, Brazil, Canada, Chile, China, Colombia, England, France, Germany, India, Italy, Japan, Mexico, Panama, Singapore, Spain and Venezuela.

device · product 3 of 3

VITROS 5600 Integrated System --- For use in the in vitro quantitative, semi-quantitative, and qualitative measurement of a variety of analytes of clinical interest, using VITROS Chemistry Products Slides, VITROS Chemistry Products MicroTip Reagents and VITROS Immunodiagnostic Products Reagents. The VITROS Software is a part of the VITROS 3600, 4600, & 5600 Systems. The VITROS 3600, 4600, & 5600 Systems share the same software base. VITROS Sample Programming consists of Sample Identification (SID), test requests, and patient demographics. Patient demographics include patient name, address, date of birth, gender, physician, and patient identification (PID).

Z-2141-2014
Recall number
Z-2141-2014
Initiated
July 07, 2014
Classification
Class II
Status
Terminated
Quantity
Domestic: 838, Foreign: 949

App-derived interpretation

Device software or design reason.device_software_design · v1.0.0
anomaly with VITROS System Software

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Sourced

Software in the Use Environment

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

Ortho Clinical Diagnostics has identified an anomaly with VITROS System Software Version 3.0 and below. Internal testing confirmed that under very specific conditions, incorrect patient demographics were associated with a sample identification (ID) number.

Code information

Serial Numbers J56000110 - J56001954 (J Numbers are analogous to Serial Numbers, for VITROS 5600 systems the J Numbers start with 5600XXXX)

Distribution pattern

Worldwide Distribution-USA (nationwide) including Puerto Rico and the countries of Australia, Brazil, Canada, Chile, China, Colombia, England, France, Germany, India, Italy, Japan, Mexico, Panama, Singapore, Spain and Venezuela.