Skip to content
Recall Observatory FDA recall evidence

Exact FDA event identifier

Event 68704

1 recalled-product record grouped only because every row carries this exact official event ID.

Open FDA event record

Event summary

Timeline bucket
June 05, 2014
Product types
Device
Classifications
Class II
Statuses
Terminated
Recalling firm wording
Siemens Medical Solutions USA, Inc

Dossier provenance

Source snapshots represented here

  • openFDA Device Enforcement Report-date coverage June 20, 2012–July 08, 2026 Known gap: openFDA describes 2004-present coverage, but the current export has no report-date records in this interval.
  • openFDA Device Recall event_date_initiated coverage June 01, 1997–June 30, 2026

Enforcement coverage uses report dates; product initiation dates can precede those bounds. Device-enrichment coverage uses FDA event_date_initiated.

How coverage and grouping work

Complete imported group

Every recalled product in this event

1 official enforcement rows

Product dates, classifications, firms, and source wording remain attached to their individual rows. Opening this dossier never hides products that did not match a prior timeline filter.

device · product 1 of 1

Artis zee and Artis zeego systems. x-ray, angiographic system

Z-2569-2014
Recall number
Z-2569-2014
Initiated
June 05, 2014
Classification
Class II
Status
Terminated
Quantity
206

App-derived interpretation

Device software or design reason.device_software_design · v1.0.0
problem with Artis zee and Artis zeego systems running software

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Sourced

Software design

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

There is a potential problem with Artis zee and Artis zeego systems running software version VC21B and being used in conjunction with the Large Display, in that under certain circumstances, the release of radiation can become blocked unnecessarily.

Code information

running software version VC21B 10094135, 10094137, 10094141, 10280959 with multiple serial numbers .

Distribution pattern

Nationwide Distribution

Field note

Send feedback

We'll only use this to respond to your feedback.