device · product 1 of 1
MK5 NX Controller--Wheelchair controller
- Recall number
- Z-2091-2014
- Initiated
- June 24, 2014
- Classification
- Class II
- Status
- Terminated
- Recalling firm
- Invacare Corporation
- Quantity
- 819
App-derived interpretation
M41 Powered Wheelchairs received through Invacare Service Parts (ISP) have controllers that were programmed in a way this is not aligned to standard factory settings. The controller during different settings such as max forward speed, max forward acceleration, max reverse speed and max turning speed can result in momentary loss of control.
Official device-enrichment evidence · Sourced
Process control
Inspect official wording and provenance
Reason for recall
M41 Powered Wheelchairs received through Invacare Service Parts (ISP) have controllers that were programmed in a way this is not aligned to standard factory settings. The controller during different settings such as max forward speed, max forward acceleration, max reverse speed and max turning speed can result in momentary loss of control.
Code information
P/N 1143839
Distribution pattern
Nationwide Distriution: CA, SC, IL, UT, TN, KS, HI, NJ, LA, TX,FL, IN, AL, AZ, MN, WI, MN, MI, AR,MA, GA, MS, NJ, NM, OK, ME, PA, NY, NC, OK, MO, WA, IA, OR,KY, MD, NE, ID, VA, WA, ND, MT, OH, NH, CT. Also shipped to Canada