Skip to content
Recall Observatory FDA recall evidence

Exact FDA event identifier

Event 68400

1 recalled-product record grouped only because every row carries this exact official event ID.

Open FDA event record

Event summary

Timeline bucket
May 27, 2014
Product types
Device
Classifications
Class III
Statuses
Terminated
Recalling firm wording
Siemens Healthcare Diagnostics, Inc.

Dossier provenance

Source snapshots represented here

  • openFDA Device Enforcement Report-date coverage June 20, 2012–July 08, 2026 Known gap: openFDA describes 2004-present coverage, but the current export has no report-date records in this interval.
  • openFDA Device Recall event_date_initiated coverage June 01, 1997–June 30, 2026

Enforcement coverage uses report dates; product initiation dates can precede those bounds. Device-enrichment coverage uses FDA event_date_initiated.

How coverage and grouping work

Complete imported group

Every recalled product in this event

1 official enforcement rows

Product dates, classifications, firms, and source wording remain attached to their individual rows. Opening this dossier never hides products that did not match a prior timeline filter.

device · product 1 of 1

Dimension¿ Integrated Chemistry Systems intended to measure a variety of analytes in human body fluids.

Z-2133-2014
Recall number
Z-2133-2014
Initiated
May 27, 2014
Classification
Class III
Status
Terminated
Quantity
15

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
There is a misassembly issue with the photometer installed on Dimension¿ Integrated Chemistry Systems instruments. Through internal investigation by Siemens Healthcare Diagnostics has confirmed that a limited number of Dimension photometers were misassembled. These photometers are performing within specifications, but may lead to a shortened life of the source lamp.

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Sourced

Employee error

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

There is a misassembly issue with the photometer installed on Dimension¿ Integrated Chemistry Systems instruments. Through internal investigation by Siemens Healthcare Diagnostics has confirmed that a limited number of Dimension photometers were misassembled. These photometers are performing within specifications, but may lead to a shortened life of the source lamp.

Code information

DE271317, DE271320, DE271323, DE271324, DE271326, DE271327, DE271328, DE271332 DE271341, DE271343, DE271325, 12252589, 2004082898, 2004082899, 224719-AX

Distribution pattern

FL LA AR IA OH OH WI IL LA KS MT

Field note

Send feedback

We'll only use this to respond to your feedback.